FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12825994
·
Received November 17, 2021
Report
- Report Number
- 9610877-2021-50388
- Event Type
- Malfunction
- Date Received
- November 17, 2021
- Date of Event
- November 5, 2021
- Report Date
- November 17, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- PMA / PMN Number
- K131855
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OBJECTIVE PRISM. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(STAIN).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1728143 | PENTAX | HDVIDEO COLONOSCOPE 3.8C 13.2T 1700L FWJ | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC-3890LI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |