BD E-Z SCRUB
Report
- Report Number
- 1710034-2021-00960
- Event Type
- Malfunction
- Date Received
- November 17, 2021
- Date of Event
- August 26, 2021
- Report Date
- December 28, 2021
- Manufacturer
- BD SANDY
- Product Code
- GEC
- UDI-DI
- 00382903716036
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
NO SAMPLES OR PHOTOS WERE PROVIDED FOR EVALUATION. AS A RESULT, BD WAS UNABLE TO VERIFY THE REPORTED ISSUE OR DETERMINE A DEFINITIVE ROOT CAUSE AT THIS TIME. REVIEW OF DEVICE HISTORY REVIEW DID NOT FIND ANY ATTRIBUTABLE DEFICIENCIES THAT WOULD LEAD TO THIS ISSUE. REVIEW OF RISK MANAGEMENT DOCUMENT SHOWS THAT SEVERITY FOR STERILE PRODUCTS FOR LEAKING SEAL IS ACCEPTABLE WITH A REMOTE OCCURRENCE FOR THIS DEFECT AND PATIENT INFECTIONS/COMPLICATIONS BEING THE POTENTIAL FAILURE MODE. POSSIBLE CAUSES OF THIS ISSUE COULD BE RESULT OF INCORRECT SET UP OR REJECTED PRODUCT BEING INADVERTENTLY INTRODUCED BACK INTO THE PRODUCT STREAM. REVIEW OF THE BATCH RECORD SHOWS THAT PRODUCTION WAS PERFORMED AS SPECIFIED AND ALL IN-PROCESS QUALITY CHECKS MET ESTABLISHED SPECIFICATIONS. REVIEW OF LOT HISTORY SHOW NO NON-CONFORMANCE EVENTS ISSUED FOR THIS LOT. ROOT CAUSE FOR THIS ISSUE CANNOT BE ESTABLISHED AT THIS TIME. ANY FUTURE COMPLAINTS WILL CONTINUE TO BE MONITORED FOR THIS LOT AND ASSESSED. H3 OTHER TEXT : SEE NARRATIVE BELOW.
PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).
IT WAS REPORTED THAT THE PACK WAS OPENED. VERBATIM: DESCRIBE THE EVENT OR PROBLEM: BD E-Z SCRUB 160 SURGICAL SCRUB BRUSH/SPONGE WITH NAIL CLEANER PACK WAS OPENED AND NOTED TO HAVE BLACK DEBRIS ON THE SPONGE.
IT WAS REPORTED THAT THE PACK WAS OPENED. DESCRIBE THE EVENT OR PROBLEM: BD E-Z SCRUB 160 SURGICAL SCRUB BRUSH/SPONGE WITH NAIL CLEANER PACK WAS OPENED AND NOTED TO HAVE BLACK DEBRIS ON THE SPONGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1729808 | BD E-Z SCRUB | IMPREGNATED PREOPERATIVE SCRUB BRUSHES (3% CHLOROXYLENOL (PCMX)) | GEC | BD SANDY | 371603 | 0328903 | 00382903716036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |