WALLACH ULTRA FREEZE
Report
- Report Number
- 1216677-2021-00266
- Event Type
- Malfunction
- Date Received
- November 16, 2021
- Date of Event
- August 18, 2021
- Report Date
- March 10, 2022
- Manufacturer
- COOPERSURGICAL, INC.
- Product Code
- GEH
- PMA / PMN Number
- K935010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION: X: REVIEW DHR. X: INSPECT RETURNED SAMPLES. ANALYSIS AND FINDINGS: COMPLAINT: (B)(4). DISTRIBUTION HISTORY: THIS COMPLAINT UNIT WAS MANUFACTURED AT CSI ON 07/31/2018 UNDER WO#: (B)(4) AND SHIPPED ON 11/28/2018. MANUFACTURING RECORD REVIEW: DHR-232774 WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW: NOT APPLICABLE. SERVICE HISTORY RECORD: THIS UNIT WAS SERVICED IN 2019 AND 2020. THE REASON FOR REPAIRS WERE DIFFERENT FROM THE CURRENT ISSUE. HISTORICAL COMPLAINT REVIEW: A REVIEW OF THE 2-YEAR COMPLAINT HISTORY SHOWED SIMILAR REPORTED COMPLAINT CONDITION. PRODUCT RECEIPT: THE COMPLAINT UNIT WAS RETURNED ON A REPAIR. HOWEVER, BASED ON LOG: 96925 THIS UNIT WAS AT CSI ON 09/01/2021. VISUAL EVALUATION: VISUAL EXAMINATION OF THE COMPLAINT UNIT REVEALED PHYSICAL DAMAGE. FUNCTIONAL EVALUATION: COMPLAINT UNIT WAS FUNCTIONALLY EVALUATED AND FOUND NOT TO FUNCTION PROPERLY. ROOT CAUSE: THE PROBABLE ROOT CAUSE WAS CUSTOMER MISUSE OR DROPPING OF THE UNIT. WAS THE COMPLAINT CONFIRMED? YES. CORRECTION AND/OR CORRECTIVE ACTION. THE COMPLAINT UNIT WAS REPAIRED AND RETURNED TO THE CUSTOMER. COOPERSURGICAL WILL CONTINUE TO TREND THIS COMPLAINT CONDITION.
COOPERSURGICAL, INC. IS CURRENTLY INVESTIGATING THE REPORTED CONDITION.
METAL SPRAY TIP BROKE OFF. CONFIRMED COMPLAINT: LUER LOCK BROKEN OFF. PER (B)(6): BROKEN TIP DOES NOT HAVE NEGATIVE EFFECT ON PATIENT/USER AS THE DEVICE WOULD NOT FREEZE SUFFICIENTLY. REPAIR ORDER LOG#: (B)(4). 1216677-2021-00266 WALLACH ULTRA FREEZE 900076 E-COMPLAINT: (B)(4).
METAL SPRAY TIP BROKE OFF. CONFIRMED COMPLAINT: LUER LOCK BROKEN OFF. PER RAY: BROKEN TIP DOES NOT HAVE NEGATIVE EFFECT ON PATIENT/USER AS THE DEVICE WOULD NOT FREEZE SUFFICIENTLY. REPAIR ORDER LOG #: 96925. WALLACH ULTRA FREEZE 900076. E-COMPLAINT- (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1723357 | WALLACH ULTRA FREEZE | WALLACH ULTRA FREEZE | GEH | COOPERSURGICAL, INC. | 900076 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |