FDA Adverse Event Malfunction Summary report: N

SERIES II LOW PROFILE MODULAR MOTOR UNIT

MDR report key: 12820520 · Received November 16, 2021

Report

Report Number
2027754-2021-00021
Event Type
Malfunction
Date Received
November 16, 2021
Report Date
March 21, 2022
Manufacturer
OSTEOMED, LLC
Product Code
GBB
UDI-DI
00845694012263
PMA / PMN Number
971692
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

H6: TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSION UPDATED TO REFLECT DEVICE EVALUATION. THE DEVICE WAS RETURNED FOR EVALUATION. NO ISSUE WAS FOUND WITH THE DEVICE, AND THE DEVICE PASSED ALL TESTS. THIS REPORT IS RELATED TO REPORT NUMBER 2027754-2021-00022 FOR THE POWER CONSOLE AND 2027754-2021-00023 FOR THE FOOTSWITCH. THE DHR WAS REVIEWED FOR THE REPORTED LOT NUMBER AND ALL INSPECTIONS PASSED PER THE DHR.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ON GOING AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ON COMPLETION OF INVESTIGATION.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ON GOING AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ON COMPLETION OF INVESTIGATION.

Description of Event or Problem · 0

ON OCTOBER 26, 2021, THE CUSTOMER CONTACTED OSTEOMED PRODUCT MANAGER BY EMAIL. PER THE EMAIL: "THE CONSOLE UNIT WILL SPONTANEOUSLY BEGIN RUNNING ON REVERSE, WITHOUT ANYONE PRESSING THE FOOT PEDAL! " THE INVOLVED PART NUMBERS ARE AS FOLLOWS: COMPLAINT #, PART #, DESCRIPTION, SERIAL #; (B)(4), 450-0084, SERIES II LOW PROFILE MODULAR MOTOR UNIT, (B)(4); (B)(4), 450-0021, POWER CONTROL CONSOLE 2 (B)(4); (B)(4), 450-0390, BMF FOOTSWITCH (B)(4).

Description of Event or Problem · 0

ON OCTOBER 26, 2021, THE CUSTOMER CONTACTED OSTEOMED PRODUCT MANAGER BY EMAIL. PER THE EMAIL: "THE CONSOLE UNIT WILL SPONTANEOUSLY BEGIN RUNNING ON REVERSE, WITHOUT ANYONE PRESSING THE FOOT PEDAL! " THE INVOLVED PART NUMBERS ARE AS FOLLOWS: COMPLAINT #, PART #, DESCRIPTION, SERIAL #; (B)(4), 450-0084, SERIES II LOW PROFILE MODULAR MOTOR UNIT, (B)(4); (B)(4), 450-0021, POWER CONTROL CONSOLE 2 (B)(4); (B)(4), 450-0390, BMF FOOTSWITCH (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1719727 SERIES II LOW PROFILE MODULAR MOTOR UNIT SERIES II LOW PROFILE MODULAR MOTOR UNIT GBB OSTEOMED, LLC 450-0084 00845694012263

Patients

Seq Age Sex Outcome Treatment
1 Unknown