MAGNUM INSTRUMENT
Report
- Report Number
- 2020394-2021-01957
- Event Type
- Malfunction
- Date Received
- November 15, 2021
- Date of Event
- October 18, 2021
- Report Date
- November 17, 2021
- Manufacturer
- BARD PERIPHERAL VASCULAR, INC.
- Product Code
- KNW
- UDI-DI
- 00801741083945
- PMA / PMN Number
- K934371
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H10: MANUFACTURING REVIEW: A DEVICE HISTORY RECORD AND MANUFACTURING REVIEW WAS NOT REQUIRED AS THE EVENT WAS DETERMINED TO BE EXPECTED AND THE INVESTIGATION DID NOT IDENTIFY ANY MANUFACTURING AND/OR SERVICE-RELATED ISSUES. INVESTIGATION SUMMARY: THE MAGNUM DRIVER WAS RECEIVED FOR EVALUATION. THE MAGNUM DRIVER WAS VISUALLY INSPECTED UPON RECEIPT AND WAS FOUND TO BE IN A GOOD CONDITION. THE DEVICE WAS FUNCTIONALLY TESTED AND PASSED THE TEST. NO OTHER ANOMALIES WERE IDENTIFIED. THEREFORE, THE INVESTIGATION IS UNCONFIRMED FOR THE REPORTED DEVICE IS FIRING BY ITSELF. THE ROOT CAUSE CANNOT BE DETERMINED AS THE PROBLEM COULD NOT BE REPRODUCED. LABELING REVIEW: A REVIEW OF PRODUCT LABELING DOCUMENTATION (E.G., PROCEDURAL INSTRUCTIONS, INDICATIONS, WARNINGS, PRECAUTIONS, CAUTIONS, POSSIBLE COMPLICATIONS, CONTRAINDICATIONS, NURSING GUIDE, AND UNIT LABEL) DID NOT FIND ANY PRODUCT LABELING INADEQUACY. H10: D4 (EXPIRY DATE: 12/2050), G3. H11: H6 (RESULT, CONCLUSION). H11: SECTION A THROUGH F ¿ THE INFORMATION PROVIDE BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
AS THE SERIAL NUMBER FOR THE DEVICE WAS PROVIDED, A REVIEW OF THE DEVICE HISTORY RECORDS IS CURRENTLY BEING PERFORMED. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INVESTIGATION OF THE REPORTED EVENT IS CURRENTLY UNDERWAY. (EXPIRY DATE: 12/2050).
IT WAS REPORTED THAT THE DEVICE IS FIRING BY ITSELF. THERE WAS NO PATIENT CONTACT.
IT WAS REPORTED THAT THE DEVICE IS FIRING BY ITSELF. THERE WAS NO PATIENT CONTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1714784 | MAGNUM INSTRUMENT | BIOPSY INSTRUMENT | KNW | BARD PERIPHERAL VASCULAR, INC. | MG1522 | 00801741083945 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |