FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12808766 · Received November 14, 2021

Report

Report Number
1221359-2021-03499
Event Type
Malfunction
Date Received
November 14, 2021
Report Date
November 20, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
04571226475027
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M162489 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000J / LOT M162489 AND TEST BASE PART NUMBER 190-430 / LOT M162489. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M162489 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. REFERENCE MFR: 1221359-2021-03498, 1221359-2021-03499, 1221359-2021-03500.

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4) ON RETAINED KIT LOT M162489 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000J / LOT M162489 AND TEST BASE PART NUMBER 190-430 / LOT M162489. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M162489 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ON A DIRECT TESTED NASOPHARYNGEAL NIPRO SWAB. CONFIRMATION TESTING ON SWABS WITH SMART GENE PCR GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1706902 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCABOROUGH M162489 04571226475027

Patients

Seq Age Sex Outcome Treatment
1 Unknown