FDA Adverse Event Malfunction Summary report: N

IMP,TSV,MCOL MG,4.7MM,10M

MDR report key: 12802464 · Received November 12, 2021

Report

Report Number
0002023141-2021-03246
Event Type
Malfunction
Date Received
November 12, 2021
Date of Event
November 1, 2021
Report Date
May 12, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019775
PMA / PMN Number
K111889
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: INVESTIGATION TYPE CODES WERE ADDED: 3331, 4109, 4111 AND 4114. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 213. H6: INVESTIGATION CONCLUSIONS CODES WERE ADDED: 67. H10: NARRATIVE/DATA WAS UPDATED. ONE (1) IMP,TSV,MCOL MG,4.7MM,10M (TSVMWB10) WAS NOT RETURNED FOR INVESTIGATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM AND LOT NUMBER ALONG WITH THE APPLICABLE DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), AND RISK FILE. THE REPORTED EVENT IS NON-VERIFIABLE. BASED ON THE EVALUATION, DEVICE MALFUNCTION COULD NOT BE VERIFIED. HOWEVER, THERE IS NO EXISTING NONCONFORMANCE/CAPA/HHE/D/IE/PRODUCT HOLDS AGAINST THE REPORTED DEVICE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCTS WERE LIKELY WITHIN SPECIFICATIONS AND LIKELY CONFORMING WHEN IT LEFT ZIMMER BIOMET. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1243376). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1243376) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: (DAMAGED THREADS) THE REPORTED EVENT COULD NOT BE RECREATED DUE TO THE NATURE OF THE DENTAL DEVICE AND EVENT AND THE COMPLAINT IS RELATED TO THE FUNCTIONAL PERFORMANCE OF THE DEVICE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE INVESTIGATION AND RISK FILE REVIEW, THE MOST LIKELY CAUSE DETERMINED FROM THE INVESTIGATION IS EXCESSIVE TORQUE ON IMPLANT. NO FURTHER INVESTIGATION AND NO IMMEDIATE CAPA/HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ADDITIONAL PMA/510(K) NUMBER ¿ K101880. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INTERNAL THREADS FOR IMPLANT AT TOOTH SITE #31 WERE DAMAGED DURING A CLINICAL PROCEDURE. THE IMPLANT IS WELL SEATED AND INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1705813 IMP,TSV,MCOL MG,4.7MM,10M DENTAL IMPLANT DZE ZIMMER DENTAL TSVMWB10 1243376 00889024019775

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male