CHLORAPREP ONE STEP HI LITE ORANGE
Report
- Report Number
- 3004932373-2021-00489
- Event Type
- Malfunction
- Date Received
- November 11, 2021
- Date of Event
- October 20, 2021
- Report Date
- December 5, 2021
- Manufacturer
- CAREFUSION 213, LLC 0113
- Product Code
- KXG
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- 501
Narratives
A SAMPLE & PHOTOGRAPHS WERE RECEIVED FOR ANALYSIS SHOWING A BLACK SPECS. ONE OF THE SOURCES OF FM BLACK SPECS IS THE PRINTER OF THE PACKAGING LIDDING. ON (B)(6)2021, THE WEEKLY PREVENTIVE MAINTENANCE WAS UPDATED TO INCLUDE A THOROUGH CLEANING OF ALL ROLLERS (ADDED IDLE ROLLERS ON THE PRINTER MACHINE, MARK ANDY EQUIPMENT). NO FURTHER ACTION WILL BE TAKEN BASED ON A NEGATIVE TREND IS NOT OBSERVED AT THIS TIME. BD WILL CONTINUE TO TRACK & TREND.
(B)(4). PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
MATERIAL NO: 930815 BATCH NO: 1193037. IT WAS REPORTED THAT THE HEALTH PROFESSIONAL FOUND BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT.¿ VERBATIM: THANK YOU FOR TAKING MY CALL AND DOCUMENTING MY QUALITY COMPLAINT. AS MENTIONED, BELOW IS THE AFFECTED PART/REFERENCE NUMBER AS WELL AS THE LOT NUMBER, AND IMAGES OF THE BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT. REF #(B)(4). LOT #1193037. PLEASE DO NOT HESITATE TO CONTACT ME SHALL YOU REQUIRE ADDITIONAL INFORMATION.
MATERIAL NO: 930815, BATCH NO: 1193037. IT WAS REPORTED THAT THE HEALTH PROFESSIONAL FOUND BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT.¿
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1698764 | CHLORAPREP ONE STEP HI LITE ORANGE | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | KXG | CAREFUSION 213, LLC 0113 | 1193037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |