FDA Adverse Event Malfunction Summary report: N

CHLORAPREP ONE STEP HI LITE ORANGE

MDR report key: 12801310 · Received November 11, 2021

Report

Report Number
3004932373-2021-00489
Event Type
Malfunction
Date Received
November 11, 2021
Date of Event
October 20, 2021
Report Date
December 5, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

A SAMPLE & PHOTOGRAPHS WERE RECEIVED FOR ANALYSIS SHOWING A BLACK SPECS. ONE OF THE SOURCES OF FM BLACK SPECS IS THE PRINTER OF THE PACKAGING LIDDING. ON (B)(6)2021, THE WEEKLY PREVENTIVE MAINTENANCE WAS UPDATED TO INCLUDE A THOROUGH CLEANING OF ALL ROLLERS (ADDED IDLE ROLLERS ON THE PRINTER MACHINE, MARK ANDY EQUIPMENT). NO FURTHER ACTION WILL BE TAKEN BASED ON A NEGATIVE TREND IS NOT OBSERVED AT THIS TIME. BD WILL CONTINUE TO TRACK & TREND.

Additional Manufacturer Narrative · 0

(B)(4). PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

MATERIAL NO: 930815 BATCH NO: 1193037. IT WAS REPORTED THAT THE HEALTH PROFESSIONAL FOUND BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT.¿ VERBATIM: THANK YOU FOR TAKING MY CALL AND DOCUMENTING MY QUALITY COMPLAINT. AS MENTIONED, BELOW IS THE AFFECTED PART/REFERENCE NUMBER AS WELL AS THE LOT NUMBER, AND IMAGES OF THE BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT. REF #(B)(4). LOT #1193037. PLEASE DO NOT HESITATE TO CONTACT ME SHALL YOU REQUIRE ADDITIONAL INFORMATION.

Description of Event or Problem · 0

MATERIAL NO: 930815, BATCH NO: 1193037. IT WAS REPORTED THAT THE HEALTH PROFESSIONAL FOUND BLACK RESIDUE UNDERNEATH THE STERILE PACKAGING THAT REMAINS INTACT.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1698764 CHLORAPREP ONE STEP HI LITE ORANGE 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXG CAREFUSION 213, LLC 0113 1193037

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other