FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12791605 · Received November 10, 2021

Report

Report Number
9610877-2021-01490
Event Type
Malfunction
Date Received
November 10, 2021
Date of Event
November 1, 2021
Report Date
November 10, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
UDI-DI
04961333237586
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: THE CAUSE IS THAT THE AIR CONTAINING MOISTURE THAT HAS ENTERED THE OBJECTIVE LENS / LED CONDENSES DUE TO TEMPERATURE CHANGES AND CAUSES FOGGING. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805.

Description of Event or Problem · 0

FOGGY, MISTY IMAGE. THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1683864 PENTAX VIDEO UPPER G.I.SCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG29-I10C 04961333237586

Patients

Seq Age Sex Outcome Treatment
1 Unknown