FDA Adverse Event Malfunction Summary report: N

THERAPLAN PLUS

MDR report key: 127867 · Received October 20, 1997

Report

Report Number
9610077-1997-00010
Event Type
Malfunction
Date Received
October 20, 1997
Date of Event
September 25, 1997
Report Date
October 16, 1997
Manufacturer
THERATRONICS INTERNATIONAL LTD.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

WHEN A PLAN IS BEING PREPARED WITH THE GANTRY AT 90 OR 270 AND COUCH AT 90 OR 270, THE SSD AND DEPTH USED FOR CALCULATION WILL CORRESPOND TO THE BEAM ENTERING ON THE FIRST SLICE. THIS WILL RESULT IN SUB-OPTIMAL CALCULATION ACCURACY IF THESE SET-UP CONDITIONS ARE USED WITH A VERTEX FIELD THAT ENTERS ON A SLICE OTHER THAN THE FIRST SLICE. THE MORE ACCURATE ENTRY-POINT SDD AND DEPTH ARE USED FOR CALCULATION FOR GANTRY ANGLES ARE OFFSET BY AT LEAST 0.5 DEGREES FROM 90 OR 270.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERAPLAN PLUS RADIATION TREATMENT PLANNING SYSTEM IWB THERATRONICS INTERNATIONAL LTD. THERAPLAN PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other