FDA Adverse Event
Malfunction
Summary report: N
THERAPLAN PLUS
MDR report key: 127867
·
Received October 20, 1997
Report
- Report Number
- 9610077-1997-00010
- Event Type
- Malfunction
- Date Received
- October 20, 1997
- Date of Event
- September 25, 1997
- Report Date
- October 16, 1997
- Manufacturer
- THERATRONICS INTERNATIONAL LTD.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
WHEN A PLAN IS BEING PREPARED WITH THE GANTRY AT 90 OR 270 AND COUCH AT 90 OR 270, THE SSD AND DEPTH USED FOR CALCULATION WILL CORRESPOND TO THE BEAM ENTERING ON THE FIRST SLICE. THIS WILL RESULT IN SUB-OPTIMAL CALCULATION ACCURACY IF THESE SET-UP CONDITIONS ARE USED WITH A VERTEX FIELD THAT ENTERS ON A SLICE OTHER THAN THE FIRST SLICE. THE MORE ACCURATE ENTRY-POINT SDD AND DEPTH ARE USED FOR CALCULATION FOR GANTRY ANGLES ARE OFFSET BY AT LEAST 0.5 DEGREES FROM 90 OR 270.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERAPLAN PLUS | RADIATION TREATMENT PLANNING SYSTEM | IWB | THERATRONICS INTERNATIONAL LTD. | THERAPLAN PLUS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |