FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1278583
·
Received January 6, 2009
Report
- Report Number
- 9611820-2008-00002
- Event Type
- Injury
- Date Received
- January 6, 2009
- Report Date
- December 10, 2008
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE DOCTOR REPORTED THAT THE PATIENT IS DOING FINE. THE DOCTOR REPLACED THE BROKEN ABUTMENT WITH A NEW BALL-HEAD POST WITHOUT ANY INCIDENT. THE PRODUCT WAS RETURNED TO SYBRON IMPLANT SOLUTIONS GMBH FOR EVALUATION. VISUAL EXAMINATION OF THE RETURNED PRODUCT INDICATED THAT THE CAUSE OF THE ABUTMENT FRACTURE WAS STRAIN THROUGH PROSTHESIS, AFTER ABUTMENT LOOSENING INTO IMPLANT-THREAD. PLEASE SEE THE ATTACHED EVALUATION REPORT. THIS IS THE FINAL REPORT.
Description of Event or Problem · 1
IN 2008, A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS GMBH THAT A PITT EASY IMPLANT BALL-HEAD POST SCREW ABUTMENT FRACTURED, DUE TO LOOSENING INTO THE THREAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Other| R |