FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1278583 · Received January 6, 2009

Report

Report Number
9611820-2008-00002
Event Type
Injury
Date Received
January 6, 2009
Report Date
December 10, 2008
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR REPORTED THAT THE PATIENT IS DOING FINE. THE DOCTOR REPLACED THE BROKEN ABUTMENT WITH A NEW BALL-HEAD POST WITHOUT ANY INCIDENT. THE PRODUCT WAS RETURNED TO SYBRON IMPLANT SOLUTIONS GMBH FOR EVALUATION. VISUAL EXAMINATION OF THE RETURNED PRODUCT INDICATED THAT THE CAUSE OF THE ABUTMENT FRACTURE WAS STRAIN THROUGH PROSTHESIS, AFTER ABUTMENT LOOSENING INTO IMPLANT-THREAD. PLEASE SEE THE ATTACHED EVALUATION REPORT. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

IN 2008, A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS GMBH THAT A PITT EASY IMPLANT BALL-HEAD POST SCREW ABUTMENT FRACTURED, DUE TO LOOSENING INTO THE THREAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 80 YR Other| R