PUMP MMT-1880L MM770G BLE MG
Report
- Report Number
- 2032227-2021-216355
- Event Type
- Malfunction
- Date Received
- November 10, 2021
- Date of Event
- July 14, 2021
- Report Date
- November 10, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000439866
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
RETAINER RING=BLACK. CUSTOMER COMPLAINED ON (B)(6) 2021 PUMP ALARMED PUMP ERROR 23. PUMP PASSED SELF TEST AND DISPLACEMENT TEST. NO PUMP ERROR 23 NOTED DURING TEST. UNIT SUCCESSFULLY DOWNLOADED TO THUMPS. VERIFIED IN PUMP ALARMED PUMP ERROR 23 AT (B)(6) 2021 14:03:11.000 IN PUMP DOWNLOADED HISTORY DUE TO INSULIN DELIVERY STOP DUE TO PUMP ALARMING PUMP ERROR 15. PUMP ERROR 15 ALARM ((B)(4)) IS FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 14:03:09.000 DUE TO A MOISTURE DAMAGE TO PCB1 BOARD AND PCB2 BOARD AS PER GLOBAL LOGIC ANALYSIS ((B)(4)). NO MOISTURE DAMAGE NOTED ON MOTOR ASSEMBLY DURING VISUAL INSPECTION. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE AND PILLOWING KEYPAD OVERLAY. THE P-CAP DOES LOCK PROPERLY. NO PUMP ERROR 23 NOTED DURING TEST. VERIFIED IN PUMP ALARMED PUMP ERROR 23 AT (B)(6) 2021 14:03:11.000 IN PUMP DOWNLOADED HISTORY DUE TO INSULIN DELIVERY STOP DUE TO PUMP ALARMING PUMP ERROR 15. PUMP ERROR 15 ALARM ((B)(4)) IS FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 14:03:09.000 DUE TO A MOISTURE DAMAGE TO PCB1 BOARD AND PCB2 BOARD AS PER GLOBAL LOGIC ANALYSIS ((B)(4)). (B)(4).
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SHOWED AN INSULIN PUMP ERROR TWICE. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ALSO COMPLETE INSULIN PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1682695 | PUMP MMT-1880L MM770G BLE MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880L | HG59ZQS | 000000763000439866 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Female |