FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM

MDR report key: 12785661 · Received November 10, 2021

Report

Report Number
0002023141-2021-03158
Event Type
Malfunction
Date Received
November 10, 2021
Date of Event
October 26, 2020
Report Date
April 25, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019515
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PREMARKET IDENTIFICATION K013227.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4) ONE TAPERED SCREW-VENT IMPLANT (TSVB11) AND MOUNT WERE RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCTS IDENTIFIED THAT THEY WERE ENGAGED AND THERE WAS BONE RESIDUE AROUND THE EXTERNAL THREADS DUE TO USAGE. FUNCTIONAL TESTING WAS PERFORMED TO DISENGAGE THE PRODUCTS. THE DEVICE COULD NOT DISENGAGE FROM EACH OTHER. DEVICE HISTORY RECORD (DHR) REVIEW FOR THE LOT (1233260) HAD REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENTS. ALL PRODUCTS WERE CONFORMING AT THE TIME THEY LEFT ZIMMER BIOMET. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1233260) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORDS: DOES NOT DISENGAGE) AND NO OTHER COMPLAINT WAS IDENTIFIED. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT (DOES NOT DISENGAGE) WAS CONFIRMED.

Description of Event or Problem · 0

DOCTOR REPORTED THAT MOUNT DID NOT DISENGAGE FROM IMPLANT THREADS. THE SURGERY WAS COMPLETED WITH ANOTHER IMPLANT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1686382 IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB11 1233260 00889024019515

Patients

Seq Age Sex Outcome Treatment
1 Unknown