FDA Adverse Event
Other
Summary report: N
EASY CAP II CO2 DETECTOR
MDR report key: 1278446
·
Received December 31, 2008
Report
- Report Number
- 2936999-2008-00648
- Event Type
- Other
- Date Received
- December 31, 2008
- Date of Event
- December 1, 2008
- Report Date
- December 19, 2008
- Manufacturer
- COVIDIEN, FORMERLY TYCO HEALTHCARE
- Product Code
- CCK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCT HAS BEEN REQUESTED FOR EVAL, IT HAS NOT BEEN REC'D. THE EASYCAP II WORKED AS EXPECTED, THE CALCIUM OXIDE PACKET IS THE ISSUE. THE REPORTER SPOKE TO MFR OF THE CALCIUM OXIDE PACKET - SUD-CHEMI - STATES PACKET WILL GENERATE HEAT ON CONTACT WITH H2O OR MOISTURE. THIS RISK IS INDICATED ON THE PACKAGING.
Description of Event or Problem · 1
IT WAS REPORTED TO COVIDIEN THAT DURING A CODE BLUE, THE CALCIUM OXIDE FRESHNESS PACKET THAT IS PACKED IN THE EASYCAP II FOIL POUCH FELL OUT OF THE POUCH AND CAME IN CONTACT WITH FLUID WHICH AS A RESULT CHEMICALLY REACTED AND BURNED THE PT. PT WAS STABILIZED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASY CAP II CO2 DETECTOR | CO2 DETECTOR | CCK | COVIDIEN, FORMERLY TYCO HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |