FDA Adverse Event Injury Summary report: N

ALERE NMP22 BLADDER CHEK 24T

MDR report key: 12783652 · Received November 9, 2021

Report

Report Number
1221359-2021-03461
Event Type
Injury
Date Received
November 9, 2021
Report Date
July 5, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
NAH
UDI-DI
10811877010569
PMA / PMN Number
P940035
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER D1200 / LOT 158575 AND DEVICE PART NUMBER F4090 / LOT 154909. QUALITY CONTROL RELEASE TESTING MET SPECIFICATIONS AND THERE WERE NO BROKEN OR UNSEALED DESICCANTS OBSERVED. THE CURRENT OVERALL INCIDENT RATE FOR BROKEN/DAMAGED/DEFECTIVE COMPONENTS FOR THIS SPECIFIC LOT BASED ON THE TOTAL DISTRIBUTED DEVICES IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED.

Additional Manufacturer Narrative · 0

CORRECTION: H10 THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER D1200 / LOT 158703 AND DEVICE PART NUMBER F4090 / LOT 154900. QUALITY CONTROL RELEASE TESTING MET SPECIFICATIONS AND THERE WERE NO BROKEN OR UNSEALED DESICCANTS OBSERVED. THE CURRENT OVERALL INCIDENT RATE FOR BROKEN/DAMAGED/DEFECTIVE COMPONENTS FOR THIS SPECIFIC LOT BASED ON THE TOTAL DISTRIBUTED DEVICES IS 16.2%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED.

Description of Event or Problem · 0

THE HOSPITAL REPORTED THAT THERE WERE AFFECTED KIT BOXES THAT HAD DAMAGED DESICCANT POUCHES. THE DESICCANT SCATTERED WHEN UNPACKING THE REAGENT BOARD AND THE PACKAGING BAG SPLASHED INTO THE EYES OF THE OPERATOR. THEY WENT TO THE OPHTHALMOLOGY DEPARTMENT FOR MEDICAL TREATMENT WITHOUT MAJOR PROBLEMS. NO OTHER SERIOUS INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1677148 ALERE NMP22 BLADDER CHEK 24T BLADDERCHECK TEST & CONTROL KIT NAH ABBOTT DIAGNOSTICS SCABOROUGH 10811877010569

Patients

Seq Age Sex Outcome Treatment
1 Male Other