FDA Adverse Event Other Summary report: N

SIMULIX RADIOTHERAPY SIMULATOR WITH CT EXTENSION

MDR report key: 127800 · Received September 25, 1997

Report

Report Number
1121753-1997-00029
Event Type
Other
Date Received
September 25, 1997
Date of Event
April 30, 1997
Report Date
September 25, 1997
Manufacturer
NUCLETRON BV
Product Code
KPQ
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
RADIOLOGIC TECHNOLOGIST

Narratives

Description of Event or Problem · 1

THREE SCANS WERE PERFORMED WITH SIMULATOR/CT EXTENSION. THE FOURTH SCAN WAS ABORTED AT 120 DEGREES WITH MESSAGE THAT TOUCHBAR WAS ACTIVATED. TECHNICIAN SHOOK AFTERSLIT WHICH FELL DOWN AND GRAZED PTS THORAX. NO INJURY OCCURRED. TWO SCREWS WERE FOUND MISSING, ONE RUSTED SCREW WAS IN PLACE. SCREWS WERE REPLACED AND LOCKTITE'D IN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIMULIX RADIOTHERAPY SIMULATOR WITH CT EXTENSION RADIOTHERAPY SIMULATOR SYSTEM KPQ NUCLETRON BV 12T NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other