FDA Adverse Event
Other
Summary report: N
SIMULIX RADIOTHERAPY SIMULATOR WITH CT EXTENSION
MDR report key: 127800
·
Received September 25, 1997
Report
- Report Number
- 1121753-1997-00029
- Event Type
- Other
- Date Received
- September 25, 1997
- Date of Event
- April 30, 1997
- Report Date
- September 25, 1997
- Manufacturer
- NUCLETRON BV
- Product Code
- KPQ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
Description of Event or Problem · 1
THREE SCANS WERE PERFORMED WITH SIMULATOR/CT EXTENSION. THE FOURTH SCAN WAS ABORTED AT 120 DEGREES WITH MESSAGE THAT TOUCHBAR WAS ACTIVATED. TECHNICIAN SHOOK AFTERSLIT WHICH FELL DOWN AND GRAZED PTS THORAX. NO INJURY OCCURRED. TWO SCREWS WERE FOUND MISSING, ONE RUSTED SCREW WAS IN PLACE. SCREWS WERE REPLACED AND LOCKTITE'D IN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIMULIX RADIOTHERAPY SIMULATOR WITH CT EXTENSION | RADIOTHERAPY SIMULATOR SYSTEM | KPQ | NUCLETRON BV | 12T | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |