FDA Adverse Event Malfunction Summary report: N

CRYO TANK

MDR report key: 12778516 · Received November 9, 2021

Report

Report Number
12778516
Event Type
Malfunction
Date Received
November 9, 2021
Date of Event
October 14, 2021
Report Date
October 27, 2021
Manufacturer
PRAXAIR DISTRIBUTION, INC.
Product Code
ECX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IL
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Description of Event or Problem · 0

AFTER CHANGING OUT THE CRYO TANK, THE CRYOABLATION MACHINE STOPPED WORKING. IT WAS FOUND THAT NITRIC OXIDE TANK CONTAINS THE WRONG GAS. THE CYRO TANK SHOULD HAVE NON-MEDICAL GRADE AIR. IT CONTAINED MEDICAL GRADE NITRIC OXIDE, WHICH IS USED BY ANESTHESIA. BOTH TANKS WERE APPROPRIATELY LABELED AND ARE SPECIFIC TO THE CRYOABLATION CONSOLE. THIS INCORRECT GAS CAUSED THE CONSOLE TO MALFUNCTION. ONCE THE CORRECT GAS FROM ANOTHER TANK WAS HOOKED UP AND TESTED, THE MACHINE WORKED FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1676247 CRYO TANK ECX PRAXAIR DISTRIBUTION, INC.

Patients

Seq Age Sex Outcome Treatment
1 Unknown