FDA Adverse Event
Malfunction
Summary report: N
CRYO TANK
MDR report key: 12778516
·
Received November 9, 2021
Report
- Report Number
- 12778516
- Event Type
- Malfunction
- Date Received
- November 9, 2021
- Date of Event
- October 14, 2021
- Report Date
- October 27, 2021
- Manufacturer
- PRAXAIR DISTRIBUTION, INC.
- Product Code
- ECX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- IL
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
Description of Event or Problem · 0
AFTER CHANGING OUT THE CRYO TANK, THE CRYOABLATION MACHINE STOPPED WORKING. IT WAS FOUND THAT NITRIC OXIDE TANK CONTAINS THE WRONG GAS. THE CYRO TANK SHOULD HAVE NON-MEDICAL GRADE AIR. IT CONTAINED MEDICAL GRADE NITRIC OXIDE, WHICH IS USED BY ANESTHESIA. BOTH TANKS WERE APPROPRIATELY LABELED AND ARE SPECIFIC TO THE CRYOABLATION CONSOLE. THIS INCORRECT GAS CAUSED THE CONSOLE TO MALFUNCTION. ONCE THE CORRECT GAS FROM ANOTHER TANK WAS HOOKED UP AND TESTED, THE MACHINE WORKED FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1676247 | CRYO TANK | ECX | PRAXAIR DISTRIBUTION, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |