FDA Adverse Event Malfunction Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 12776329 · Received November 8, 2021

Report

Report Number
1221359-2021-03442
Event Type
Malfunction
Date Received
November 8, 2021
Date of Event
October 5, 2021
Report Date
March 25, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
04571226475027
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT LOT M162309 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000J / LOT M162309, TEST BASE PART NUMBER 190-430 / LOT M162309. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M162309 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. MFR REF REPORTS: 1221359-2021-03443.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWO (2) CONFLICTING RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021. THIS MFR. REPORT PATIENT ONE(1) AND IS ONE (1) OF TWO (2). THE CUSTOMER REPORTED A CONFLICTING RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWO (2) UNCONFIRMED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON DIFFERENT DATES. THIS MFR. REPORT ADDRESSES DATE (B)(6) 2021 AND IS ONE (1) OF TWO (2). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1672290 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M162309 04571226475027

Patients

Seq Age Sex Outcome Treatment
1 Unknown