FDA Adverse Event Malfunction Summary report: N

BD E-Z SCRUB¿

MDR report key: 12775290 · Received November 8, 2021

Report

Report Number
1710034-2021-00961
Event Type
Malfunction
Date Received
November 8, 2021
Date of Event
August 26, 2021
Report Date
December 28, 2021
Manufacturer
BD SANDY
Product Code
GEC
UDI-DI
00382903716036
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4)

Additional Manufacturer Narrative · 0

SAMPLE OR PHOTO NOT SUBMITTED FOR REVIEW. BATCH RECORD REVIEW SHOWS THAT ALL ATTRIBUTED DATA PASSED QUALITY CHECKS THROUGHOUT THE PRODUCTION OF THE LOT, AND NO DEVIATIONS OR DISCREPANCIES WERE NOTED THAT WOULD CONTRIBUTE OR CAUSE THIS ISSUE. REVIEW OF RISK MANAGEMENT DOCUMENT SHOWS THAT SEVERITY RATING IS MINOR FOR THIS DEFECT. THIS IS THE FIRST COMPLIANT REPORTED FOR THIS LOT. SIMILAR COMPLAINT (B)(4) IDENTIFIED THE FOREIGN MATTER TO LIKELY BE PART OF THE FOAM SPONGE THAT WAS BURNT AND NOT A FOREIGN OBJECT TO THE SPONGE. DUE TO LOW COMPLAINT HISTORY FOR THIS LOT, DATA FROM IN-PROCESS QUALITY CHECKS DURING PRODUCTION, AND QUALITY CONTROL LABORATORY DATA, NO FURTHER ACTION IS RECOMMENDED AT THIS TIME. COMPLAINTS WILL CONTINUE TO BE TRENDED AND EVALUATED FOR FURTHER ACTION, IF NEEDED. NO FURTHER ACTIONS REQUIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PACK WAS NOTED TO HAVE BLACK DEBRIS ON THE SPONGE. VERBATIM: DESCRIBE THE EVENT OR PROBLEM: BD E-Z SCRUB 160 SURGICAL SCRUB BRUSH/SPONGE WITH NAIL CLEANER PACK WAS OPENED AND NOTED TO HAVE BLACK DEBRIS ON THE SPONGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PACK WAS NOTED TO HAVE BLACK DEBRIS ON THE SPONGE. VERBATIM: DESCRIBE THE EVENT OR PROBLEM: BD E-Z SCRUB 160 SURGICAL SCRUB BRUSH/SPONGE WITH NAIL CLEANER PACK WAS OPENED AND NOTED TO HAVE BLACK DEBRIS ON THE SPONGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1674280 BD E-Z SCRUB¿ IMPREGNATED PREOPERATIVE SCRUB BRUSHES (3% CHLOROXYLENOL (PCMX) GEC BD SANDY 371603 0328903 00382903716036

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other