FDA Adverse Event Other Summary report: N

60" ATLAS TRANSFER CARRIAGE

MDR report key: 1277511 · Received November 14, 2008

Report

Report Number
2515984-2008-00015
Event Type
Other
Date Received
November 14, 2008
Date of Event
October 2, 2008
Report Date
January 14, 2008
Manufacturer
STERIS MEXICO S. DE R.L. DE C.V
Product Code
FLE
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN VISITED THE HOSPITAL AND DETERMINED THE WHEEL CAME LOOSE FROM THE CART AND, CAUSED IT TO TIP OVER. THE CART IS TWO YEARS OLD. IT WAS MANUFACTURED ON (B) (6) 2006. NO SERIOUS INJURY WAS SUSTAINED AS A RESULT OF THE INCIDENT.

Description of Event or Problem · 1

WHEN MOVING A LOADED TRANSFER CART, A WHEEL CAME LOOSE CAUSING THE CART TO TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 60" ATLAS TRANSFER CARRIAGE TRANSFER CARRIAGE FLE STERIS MEXICO S. DE R.L. DE C.V 4011199 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention