FDA Adverse Event Injury Summary report: N

TOTALLIFT -EPL

MDR report key: 1277305 · Received December 31, 2008

Report

Report Number
MW5009504
Event Type
Injury
Date Received
December 31, 2008
Date of Event
December 7, 2008
Report Date
December 31, 2008
Manufacturer
WY-EAST MEDICAL CORP.
Product Code
FNG
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT WAS BEING LIFTED FROM THE FLOOR WITH THE TOTALLIFT -EPL. THE STRAP GOT CAUGHT UNDER THE LEG OF THE EQUIPMENT. THE PT EXPERIENCED SOME RESPIRATORY DISTRESS WHILE IN THE LIFT. THE EMERGENCY RELEASE DOWN BUTTON FAILED TO WORK. STAFF HAD TO REMOVE THE SLING TO GET PT OUT OF LIFT. THE PT THEN EXPERIENCED A RESPIRATORY AND CARDIAC ARREST. RESUSCITATION WAS SUCCESSFUL AT THAT TIME. INITIAL CHECKING OF THE EQUIPMENT COULD NOT DUPLICATE THE FAILURE. IN LATE 2008, OUR BIOMED DEPT PERFORMED A SECOND TEST ON THE EQUIPMENT AND AT THAT TIME THE EMERGENCY RELEASE DOWN BUTTON DID NOT FUNCTION. DATES OF USE: 2006 - 2008. DIAGNOSIS OR REASON FOR USE: PT LIFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALLIFT -EPL PATIENT LIFT FNG WY-EAST MEDICAL CORP. VL-620

Patients

Seq Age Sex Outcome Treatment
1 90 YR Life Threatening