HEATER COOLER UNIT
Report
- Report Number
- 8010762-2021-00592
- Event Type
- Malfunction
- Date Received
- November 8, 2021
- Date of Event
- October 22, 2021
- Report Date
- February 21, 2022
- Manufacturer
- MAQUET CARDIOPULMONARY GMBH
- Product Code
- DWC
- PMA / PMN Number
- K031544
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IT WAS REPORTED THAT THE HCU 30 TURNED OFF DURING PATIENT TREATMENT WITH A BURNING SMELL. THE DEVICE HAS BEEN EXCHANGED DURING PATIENT TREATMENT. NO HARM TO ANY PERSON HAS BEEN REPORTED. THE GETINGE FIELD SERVICE TECHNICIAN (FST) WAS ON SITE FOR FURTHER INVESTIGATION. THE FST CONFIRMED THE FAILURE. ACCORDING TO THE FST THE DEVICE DID NOT HAVE LEAKAGE OR ERROR MESSAGE DISPLAYED. THE LAST PREVENTIVE MAINTENANCE WAS PERFORMED ON DECEMBER 2020. THE FST HAS CONFIRMED THAT A SHORT CIRCUIT WAS CAUSED BY A BROKEN WIRE JACKET IN THE POWER SUPPLY UNIT. THE SHORT CIRCUIT HAS CAUSED THE DEVICE TO TURN OFF. THE MOST PROBABLE ROOT CAUSE FOR THE WIRE JACKET TO BE BROKEN IS THE AGING OF THE DEVICE. IT HAS BEEN DECIDED TO NOT REPAIR DUE TO AGING OF THE DEVICE. THE HCU 30 HAS BEEN PHASED OUT SINCE THE END OF 2010. THE TECHNICAL SUPPORT, IN-HOUSE REPAIRS, AS WELL AS SPARE PARTS SUPPLIES HAVE BEEN DISCONTINUED IN 2017. THE DEVICE IS TAKEN OUT OF USE. THE REVIEW OF THE NON-CONFORMITIES HAS BEEN PERFORMED ON 2021-10-26 FOR THE PERIOD OF 2004-11-04 TO 2021-10-23. IT DOES NOT SHOW ANY NON-CONFORMITY IN REGARD TO THE REPORTED PRODUCT AND FAILURE. THERE IS NO INDICATION ON MANUFACTURING ISSUES OCCURRED DURING THIS TIME, THUS PRODUCTION RELATED INFLUENCES ARE UNLIKELY TO HAVE CONTRIBUTED TO THE REPORTED EVENT. THE PRODUCT IN QUESTION WAS PRODUCED IN 2004-11-04. IF SIGNIFICANT INFORMATION BECOMES AVAILABLE, THE COMPLAINT WILL BE REOPENED AND NECESSARY STEPS WILL BE INITIATED. THE OCCURRENCE RATE RELATED TO THE REPORTED ISSUE IS CURRENTLY BEING MONITORED AS PART OF MAQUET CARDIOPULMONARY¿ S TRENDING PROGRAM. ADDITIONAL INVESTIGATIONS, OR CORRECTIVE ACTIONS, WILL BE IMPLEMENTED IN THE EVENT OF ADVERSE TRENDING.
FURTHER INVESTIGATION IS ON-GOING. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
COMPLAINT (B)(4).
COMPLAINT #(B)(4). IT WAS REPORTED THAT THE HCU 30 HAD A BURNING SMELL AND IT TURNED OFF DURING CLINICAL USE. THE MACHINE WAS REPLACED DURING PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1666015 | HEATER COOLER UNIT | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | MAQUET CARDIOPULMONARY GMBH | HCU 30 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose |