FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12757224 · Received November 7, 2021

Report

Report Number
9610877-2021-50267
Event Type
Malfunction
Date Received
November 7, 2021
Date of Event
October 28, 2021
Report Date
November 5, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FNL-10RP3 IS AVAILABLE IN THE USA WITH A 510K NUMBER K951196. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB IMAGE FAILURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE LEAKAGE OCCURRED IN THE CFB UNIT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1660905 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-10RBS

Patients

Seq Age Sex Outcome Treatment
1 Unknown