FDA Adverse Event Malfunction Summary report: N

HD 3CMOS AUTOCLAVABLE CAMERA HEAD

MDR report key: 12757025 · Received November 7, 2021

Report

Report Number
8010047-2021-14125
Event Type
Malfunction
Date Received
November 7, 2021
Date of Event
October 21, 2021
Report Date
January 21, 2022
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
OWN
PMA / PMN Number
K200542
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS AUSTRALIA & NEW ZEALAND (OAZ). OMSC CHECKED THE DEVICE HISTORY RECORD OF THE SUBJECT DEVICE, THERE WAS NO IRREGULARITY FOUND. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER, BASED ON THE INFORMATION PROVIDED BY OAZ, THE CABLE UNIT HAD BEEN REPLACED, AND OMSC SURMISED THAT A COMMUNICATION FAILURE CAUSED BY A CABLE UNIT FAILURE RESULTED IN COMMUNICATION ERROR E226. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) CHECKED THE SUBJECT DEVICE AND DUPLICATED THE REPORTED PHENOMENON. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 0

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY THE USER THAT DURING THE PROCEDURE, THE SCOPE COMMUNICATION ERROR E226 WAS DISPLAYED ON THE MONITOR. ERROR CODE E226 MEANS THAT THE COMMUNICATION BETWEEN THE ENDOSCOPE AND VIDEO SYSTEM CENTER HAS FAILED. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1659362 HD 3CMOS AUTOCLAVABLE CAMERA HEAD AUTOCLAVABLE CAMERA HEAD OWN OLYMPUS MEDICAL SYSTEMS CORP. CH-S200-XZ-EA

Patients

Seq Age Sex Outcome Treatment
1 Unknown