FDA Adverse Event Malfunction Summary report: N

BD INTIMA II¿ IV CATHETER PRN ADAPTER

MDR report key: 12755990 · Received November 6, 2021

Report

Report Number
3014704491-2021-00240
Event Type
Malfunction
Date Received
November 6, 2021
Date of Event
September 19, 2021
Report Date
October 14, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED, A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9050858. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA II¿ IV CATHETER PRN ADAPTER LEAKED FROM THE END CAP DURING THE INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "THE PATIENT UNDERWENT LAPAROSCOPIC PARTIAL RESECTION OF THE LEFT KIDNEY MASS, AND POSTOPERATIVE SYMPTOMATIC FLUID REPLENISHMENT WAS FOUND. WHEN THE INDWELLING NEEDLE WAS USED TO DRAIN THE FLUID DURING INTRAVENOUS INFUSION AT 9:00 ON (B)(6) 2021, THE END CAP LEAKAGE WAS FOUND, WHICH WAS IMMEDIATELY REPLACED WITHOUT INJURY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1653938 BD INTIMA II¿ IV CATHETER PRN ADAPTER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1051529

Patients

Seq Age Sex Outcome Treatment
1 Unknown