FDA Adverse Event Malfunction Summary report: N

EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE

MDR report key: 12750338 · Received November 4, 2021

Report

Report Number
8010047-2021-14068
Event Type
Malfunction
Date Received
November 4, 2021
Report Date
November 4, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE MANUFACTURING RECORD WAS REVIEWED AND FOUND NO IRREGULARITIES. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED, BUT THERE WAS A POSSIBILITY OF INSUFFICIENT REPROCESSING OR HANDLING PROBLEM.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING THE INCOMING INSPECTION FOR REPAIR AT OLYMPUS SERVICE OPERATION REPAIR CENTER (SORC) ON (B)(6) 2021, IT WAS FOUND THAT THE NOZZLE HAD A FOREIGN MATERIAL AND WAS CLOGGED. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1653224 EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS OLYMPUS MEDICAL SYSTEMS CORP. GIF-HQ290

Patients

Seq Age Sex Outcome Treatment
1