FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ PEN NEEDLE

MDR report key: 12745561 · Received November 3, 2021

Report

Report Number
9616656-2021-01365
Event Type
Malfunction
Date Received
November 3, 2021
Date of Event
October 5, 2021
Report Date
December 22, 2021
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 2021-12-16. H6: INVESTIGATION SUMMARY 81 SEALED 31G X 5MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 0162448, CAT. NO. 320522. A CLOG TEST AS PER TP700483 WAS CARRIED OUT ON THIRTY SAMPLES AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED BD ULTRA-FINE¿ PEN NEEDLE WAS UNABLE TO DELIVER INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "HER CUSTOMER SAYS THAT NO INSULIN COMES THROUGH..."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED BD ULTRA-FINE¿ PEN NEEDLE WAS UNABLE TO DELIVER INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "HER CUSTOMER SAYS THAT NO INSULIN COMES THROUGH..."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1647001 BD ULTRA-FINE¿ PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 0162448

Patients

Seq Age Sex Outcome Treatment
1 Unknown