FDA Adverse Event Malfunction Summary report: N

MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 12743721 · Received November 3, 2021

Report

Report Number
2032227-2021-214121
Event Type
Malfunction
Date Received
November 3, 2021
Date of Event
October 25, 2021
Report Date
April 5, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000414344
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

RETAINER RING=BLACK. CUSTOMER COMPLAINED ON 10/25/2021 THE DEVICE ALARMED PUMP ERROR 49 AND HAS A CONSTANT TONE. DEVICE PASSED DISPLACEMENT TEST, SELF TEST, ACTIVE CURRENT MEASUREMENT AND SLEEP CURRENT MEASUREMENT. DEVICE WAS MONITORED. NO PUMP ERROR 49 OR CONSTANT TONE NOTED DURING TESTING. UNIT SUCCESSFULLY DOWNLOADED TO THUMP. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO POWER ANOMALIES NOTED DURING TESTING OR IN THE DEVICE TRACE DOWNLOAD. CONFIRMED PUMP ERROR 49 IN THE HISTORY DOWNLOAD ON 10/25/2021 13:29:54.000 DUE TO CORRUPTED HISTORY POINTERS, PROBLEM ISOLATED TO THE PCB2 BOARD. THE ELECTRONIC ASSEMBLY, BATTERY TUBE AND BATTERY CAP WERE INSPECTED AND NO ANOMALIES NOTED. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE AND PILLOWING KEYPAD OVERLAY. THE P-CAP/RESERVOIR DOES LOCK PROPERLY. NO PUMP ERROR 49 OR CONSTANT TONE NOTED DURING TESTING. HOWEVER, CONFIRMED PUMP ERROR 49 IN THE HISTORY DOWNLOAD ON 10/25/2021 13:29:54.000 DUE TO CORRUPTED HISTORY POINTERS, PROBLEM ISOLATED TO THE PCB2 BOARD. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A INSULIN PUMP ERROR ALARM. THE CUSTOMER STATED THEY WERE NOT ABLE TO CLEAR THE ALARM. CUSTOMER STATED THAT INSULIN PUMP WAS MAKING BEEPING SOUND BUT THERE WAS NO MESSAGE ON THE SCREEN. CUSTOMER STATED THAT THEY HEARD BEEPING SOUND AND INSULIN PUMP WAS NOT SHOWING ANYTHING ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1640599 MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1880 HG5868GZZ 000000763000414344

Patients

Seq Age Sex Outcome Treatment
1 80 YR Male