CODMAN BACTISEAL CATHETER KIT
Report
- Report Number
- 3013886523-2021-00463
- Event Type
- Injury
- Date Received
- November 2, 2021
- Date of Event
- October 18, 2021
- Report Date
- February 16, 2022
- Manufacturer
- INTEGRA LIFESCIENCES SWITZERLAND SAR
- Product Code
- JXG
- PMA / PMN Number
- K102589
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE BACTISEAL CATHETER (823072) WAS NOT RETURNED FOR EVALUATION (AS PER CUSTOMER, PRODUCT NOT AVAILABLE). LOT NUMBER INFORMATION HAS NOT BEEN PROVIDED; THEREFORE, AN EVALUATION OF THE DEVICE COULD NOT BE PERFORMED, AND DEVICE HISTORY RECORD (DHR) COULD NOT BE REVIEWED. THE ROOT CAUSE(S) OF THE REPORTED ISSUE COULD NOT BE DETERMINED. HOWEVER, THE POSSIBLE ROOT CAUSE FOR THE ISSUE REPORTED BY THE CUSTOMER COULD BE DUE TO THE DEVICE CONTAINING LATEX¿ AND POTENTIAL EFFECTS OF FAILURE CAN BE ¿PATIENT REACTION TO LATEX". IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED. UNIQUE DEVICE IDENTIFICATION (UDI):(B)(4).
AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
A PHYSICIAN REPORTED A BACTISEAL CATHETER (ID823072) WAS IMPLANTED IN (B)(6) 2021 AND WAS USED WITH 823112 (SERIAL;UNK). AN INFLAMMATORY REACTION WAS REPORTED THAT CAUSED SWELLING AND REDNESS ALONG THE CATHETER. BASED ON INFORMATION PROVIDED, CATHETER FAILURE IS UNKNOWN AND NO TREATMENT WAS GIVEN FOR THE INFLAMMATORY REACTION. THE PATIENT'S CONDITION IS STABLE AND THE SWELLING HAS DISAPPEARED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1637129 | CODMAN BACTISEAL CATHETER KIT | BACTISEAL CATHETER KIT | JXG | INTEGRA LIFESCIENCES SWITZERLAND SAR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | Unknown |