XPS HANDPIECE - VISAO
Report
- Report Number
- 1045254-2021-00606
- Event Type
- Malfunction
- Date Received
- November 2, 2021
- Report Date
- November 2, 2021
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- ERL
- UDI-DI
- 00885074107332
- PMA / PMN Number
- K011321
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 3334625, SERIAL/LOT #: N/A. PMA / 510(K) #: K983224. PRODUCT ID: 3334800 ANALYSIS FOUND THE HANDPIECE NOISY AND OVERHEATING. AFTER ONE MINUTE THE PRE-REPAIR TEMPERATURES WERE REAR 50.9 DEG C, MIDDLE 60.2 DEG C AND NOSE 51.4 DEGC. PRODUCT ID 3334625 DURING INSPECTION, IT WAS CONFIRMED THAT THE BEARING HAD DETERIORATED. THERE WAS OVERHEATING. AFTER ONE MINUTE THE PRE-REPAIR TEMPERATURE WAS 49.8 DEGC. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT THE DEVICES GENERATED HEAT DURING INTRA-OP. THERE IS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED THAT THE VISAO DRILL OVERHEATED WITH THE 3334625 BUR GUARD AT 49.8C AT ONE MINUTE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1635912 | XPS HANDPIECE - VISAO | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I | ERL | MEDTRONIC XOMED INC. | 3334800 | 73520500 | 00885074107332 |