FDA Adverse Event Malfunction Summary report: N

XPS HANDPIECE - VISAO

MDR report key: 12738015 · Received November 2, 2021

Report

Report Number
1045254-2021-00606
Event Type
Malfunction
Date Received
November 2, 2021
Report Date
November 2, 2021
Manufacturer
MEDTRONIC XOMED INC.
Product Code
ERL
UDI-DI
00885074107332
PMA / PMN Number
K011321
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 3334625, SERIAL/LOT #: N/A. PMA / 510(K) #: K983224. PRODUCT ID: 3334800 ANALYSIS FOUND THE HANDPIECE NOISY AND OVERHEATING. AFTER ONE MINUTE THE PRE-REPAIR TEMPERATURES WERE REAR 50.9 DEG C, MIDDLE 60.2 DEG C AND NOSE 51.4 DEGC. PRODUCT ID 3334625 DURING INSPECTION, IT WAS CONFIRMED THAT THE BEARING HAD DETERIORATED. THERE WAS OVERHEATING. AFTER ONE MINUTE THE PRE-REPAIR TEMPERATURE WAS 49.8 DEGC. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT THE DEVICES GENERATED HEAT DURING INTRA-OP. THERE IS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED THAT THE VISAO DRILL OVERHEATED WITH THE 3334625 BUR GUARD AT 49.8C AT ONE MINUTE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1635912 XPS HANDPIECE - VISAO DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I ERL MEDTRONIC XOMED INC. 3334800 73520500 00885074107332

Patients

Seq Age Sex Outcome Treatment
1