FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 12737610 · Received November 2, 2021

Report

Report Number
1810909-2021-00376
Event Type
Malfunction
Date Received
November 2, 2021
Date of Event
October 13, 2021
Report Date
October 13, 2021
Manufacturer
ASCENSIA DIABETES CARE US INC.
Product Code
NBW
PMA / PMN Number
K111268
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER RETURNED THE SUSPECTED CONTOUR XT METER AND CONTOUR NEXT TEST STRIPS FROM LOT # 0JPEG16A FOR EVALUATION. THE RETURNED VIAL OF TEST STRIPS CONTAINED ONLY THREE TEST STRIPS WHICH WERE USED FOR THE TESTING OF THE RELATED ISSUE WITH THE CONTOUR NEXT LINK 2.4 METER. THEREFORE, THE RETURNED METER WAS TESTED WITH THE IN-HOUSE CONTOUR NEXT TEST STRIPS FROM LOT # 0JPEG16A USING A BLOOD SAMPLE, WHICH GAVE A SATISFACTORY PERFORMANCE.

Additional Manufacturer Narrative · 0

THE INITIAL REPORT INDICATED THAT THE DEVICE INVOLVED IN THE EVENT OCCURRENCE WAS THE CONTOUR NEXT (CONNECTED) METER. BASED ON THE ADDITIONAL INFORMATION RECEIVED, THE DEVICE INVOLVED IN THE EVENT OCCURRENCE WAS THE CONTOUR XT METER. SECTIONS B5 AND G4 HAVE BEEN UPDATED WITH THE CORRECT INFORMATION.

Description of Event or Problem · 0

AN ADVOCATE FROM GERMANY CALLED TO REPORT THAT WITHIN 10 MINUTES HIS SON OBTAINED BLOOD GLUCOSE READINGS OF 13 MG/DL AND 222 MG/DL WITH THE CONTOUR XT METER.

Additional Manufacturer Narrative · 1

THE PATIENT/FAMILY WAS THE INITIAL REPORTER, SO PERSONAL INFORMATION WAS NOT ENTERED. NO INFORMATION WAS CAPTURED AS THE CUSTOMER'S AGE AND WEIGHT WERE NOT PROVIDED. THE MODEL # WAS NOT PROVIDED. THE CONTOUR NEXT (CONNECTED) METER IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE CONTOUR NEXT ONE METER WITH THE PRODUCT CODE NBW AND 510 (K) # K160682, WHICH IS MARKETED IN THE UNITED STATES. THIS EVENT IS RELATED TO MDR # 1810909-2021-00375.

Description of Event or Problem · 1

AN ADVOCATE FROM GERMANY CALLED TO REPORT THAT WITHIN 10 MINUTES HIS SON OBTAINED BLOOD GLUCOSE READINGS OF 13 MG/DL AND 222 MG/DL WITH THE CONTOUR NEXT (CONNECTED) METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE ADVOCATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1639879 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW ASCENSIA DIABETES CARE US INC. 0JPEG16A

Patients

Seq Age Sex Outcome Treatment
1 Male