FDA Adverse Event Malfunction Summary report: Y

COOK MEDICAL INC.

MDR report key: 12736325 · Received November 2, 2021

Report

Report Number
3008988055-2021-00002
Event Type
Malfunction
Date Received
November 2, 2021
Date of Event
September 22, 2021
Report Date
September 23, 2021
Manufacturer
HERAEUS MEDICAL COMPONENTS
Product Code
DQX
UDI-DI
00816840020777
PMA / PMN Number
K191502
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DATE OF THIS REPORT: 23-SEPT-2021

Additional Manufacturer Narrative · 1

REPORT DELAYED DUE TO ESUBMITTER SETUP ISSUES; COMMUNICATED WITH FDA REGARDING THIS DELAY (B)(6) 2021.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT UNDERWENT AN INTERVENTION SFA CTO IN WHICH THE APPROACH AUTHORITY WORKHORSE MICROWIRE GUIDE, ,018 WIRE DIAMETER, 300CM LENGTH, 6.5CM POLYMER LENGTH, 8CM TAPER LENGTH, G57804, WAS USED. TIBIAL ACCESS, ATTEMPTING TO CROSS CALCIFIED CTO OF SFA WITH AUTHORITY WIRE THROUGH 018 ANGLED CXI. PROLAPSED TIP TRYING TO PUSH THROUGH CTO, WIRE AND CATHETER ENTERED SUBINTIMAL SPACE. AFTER A LITTLE BIT OF MANEUVERING A PORTION OF THE TIP OF THE WIRE BROKE. DISTAL TIP OF THE WIRE BROKE OFF AND WEDGED INTO THE SUBINTIMAL SPACE, MD [THE PHYSICIAN] DECIDED TO LEAVE BROKEN TIP AND NOT TRY TO RETRIEVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1634892 COOK MEDICAL INC. APPROACH AUTHORITY WORKHORSE MICROWIRE GUIDE DQX HERAEUS MEDICAL COMPONENTS AUTH-18-300 9007497392 00816840020777

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male Other