CRYOSPRAY ABLATION SYSTEM
Report
- Report Number
- 3004534508-2008-00011
- Event Type
- Death
- Date Received
- December 22, 2008
- Date of Event
- November 24, 2008
- Report Date
- December 22, 2008
- Manufacturer
- CSA MEDICAL INC
- Product Code
- GEH
- PMA / PMN Number
- K060555
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE ATTENDING PHYSICIAN DISCONNECTED THE VENTILATOR PRIOR TO THE FIRST CRYOSPRAY TREATMENT. AFTER RECONNECTING THE VENTILATOR, THE PEAK PRESSURE WENT UP, SUGGESTING SOME TYPE OF INTRATHORACIC MALFUNCTION. NONETHELESS, THE PHYSICIAN CONTINUED WITH A SECOND CRYOSPRAY TREATMENT. AFTER RECONNECTING THE VENTILATOR AFTER THE SECOND TREATMENT, THE PEAK PRESSURES WENT UP AGAIN, SUGGESTING CONTINUED INTRATHORACIC MALFUNCTION. THE PATIENT THEN DEVELOPED BRADYCARDIA. AN X-RAY REVEALED BILATERAL PNEUMOTHORACIES. BILATERAL CHEST TUBES WERE INSERTED WITH BRIEF IMPROVEMENT IN GAS EXCHANGE BEFORE PT ULTIMATELY PASSED AWAY. CSA MEDICAL PERSONNEL EVALUATED THE CONSOLE AND FOUND IT TO BE OPERATING WITHIN SPECIFICATIONS. A MEDICAL ADVISOR TO CSA MEDICAL REVIEWED THIS SITUATION WITH THE ATTENDING PHYSICIAN. THE PHYSICIAN INFORMED CSA MEDICAL THAT HE UNDERTOOK THIS TREATMENT AS AN EXTREME LIFE SUSTAINING MEASURE. THE OPINION OF THE MEDICAL ADVISOR IS THAT THERE WERE MULTIPLE FACTORS CONTRIBUTING TO THIS EVENT, INCLUDING BUT NOT LIMITED TO THE PT'S ADVANCED AGE, EXTENSIVE AMOUNTS OF NECROTIC DEBRIS AND TISSUE IN THE AIRWAYS, ADVANCED STAGE OF LUNG CANCER, RESPIRATORY FAILURE, CRYOSPRAY THERAPY, AND HIGH BASELINE VENTILATOR PRESSURES.
PT WAS CRITICALLY ILL SECONDARY TO ADVANCED LUNG CANCER AND RESPIRATORY FAILURE. PT WAS ALSO ON A MECHANICAL VENTILATOR DUE TO THEIR TERMINAL ILLNESS AND RESPIRATORY FAILURE. BRONCHOSCOPIC EXAMINATION REVEALED EXTENSIVE AMOUNTS OF NECROTIC DEBRIS AND TISSUE IN THE AIRWAYS WITH ENDOLUMINAL OBSTRUCTION. PHYSICIAN USED SPRAY CRYOTHERAPY TO IMPROVE OPENING OF THE AIRWAYS. FOLLOWING TREATMENT AND RE-VENTILATION, THE PT DEVELOPED BRADYCARDIA AND BILATERAL PNEUMOTHORACIES. DESPITE AGGRESSIVE RESUSCITATION EFFORTS, THE PT EXPIRED. EXACT CAUSE OF DEATH IS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CRYOSPRAY ABLATION SYSTEM | CRYOSPRAY | GEH | CSA MEDICAL INC | CC2-NAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Death |