ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-03354
- Event Type
- Malfunction
- Date Received
- November 1, 2021
- Date of Event
- September 7, 2021
- Report Date
- November 24, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. REFERENCE MFR REPORT: 1221359-2021-03355. 1221359-2021-03356.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M162595 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: M162595 , TEST BASE PART NUMBER 190-430 / LOT: M162595 . THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M162595 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, AS THE LOGFILES WERE NOT PROVIDED; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION.
THE CUSTOMER REPORTED FOUR (4) FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON 2 DAYS WITH 3 LOTS. THIS MFR. REPORT IS ONE (1) OF THREE (3) REPORTS FOR ONE (1) FALSE POSITIVE RESULT ON (B)(6) 2021 WITH LOT NUMBER M162595. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT ON A NASAL SWAB WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021. CONFIRMATION TESTING WAS PERFORMED ON (B)(6) 2021 WITH CORELAB AND GENERATED NEGATIVE RESULTS (CT VALUES NOT PROVIDED). ALTHOUGH REQUESTED, ADDITIONAL INFORMATION INCLUDING PATIENT TREATMENT AND OUTCOME WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1626362 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | M162595 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |