FDA Adverse Event
Malfunction
Summary report: N
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
MDR report key: 12726701
·
Received November 1, 2021
Report
- Report Number
- 8010047-2021-13872
- Event Type
- Malfunction
- Date Received
- November 1, 2021
- Report Date
- November 1, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE MANUFACTURING RECORD WAS REVIEWED AND FOUND NO IRREGULARITIES. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED, BUT THERE WAS A POSSIBILITY OF INSUFFICIENT CLEANING.
Description of Event or Problem · 1
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING THE INCOMING INSPECTION FOR REPAIR AT OLYMPUS SERVICE OPERATION REPAIR CENTER (SORC) ON (B)(6) 2021, IT WAS FOUND THAT THERE WAS A FOREIGN MATERIAL IN THE NOZZLE DUE TO INSUFFICIENT CLEANING. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1631504 | EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE | GASTROINTESTINAL VIDEOSCOPE | FDS | OLYMPUS MEDICAL SYSTEMS CORP. | GIF-XP290N |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |