HEALON5 PRO OVD
Report
- Report Number
- 2020664-2021-07964
- Event Type
- Injury
- Date Received
- October 30, 2021
- Date of Event
- October 4, 2021
- Report Date
- October 29, 2021
- Manufacturer
- AMO UPPSALA AB
- Product Code
- LZP
- UDI-DI
- 05050474701632
- PMA / PMN Number
- P810031
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
PT INFORMATION UNKNOWN NOT PROVIDED. IF IMPLANTED, GIVE DATE: NOT APPLICABLE, THE HEALON IS NOT AN IMPLANTED PRODUCT. IF EXPLANTED, GIVE DATE: NOT APPLICABLE, THE HEALON IS NOT AN IMPLANTED PRODUCT. CONCOMITANT MEDICAL PRODUCTS: SIGNATURE PRO MACHINES AND WHITESTAR HANDPIECES. TELEPHONE NUMBER: (B)(6). DEVICE EVALUATION: AT THE TIME OF THE INVESTIGATION, PRODUCT WAS NOT AVAILABLE FOR EVALUATION, THEREFORE NO TESTING COULD BE PERFORMED. THE REPORTED EVENT CANNOT BE CONFIRMED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE PRODUCT WERE REVIEWED. FOR THIS LOT, ALL DEVICES MEET MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS AT THE TIME OF PRODUCT RELEASE. A SEARCH FOR RELATED COMPLAINTS FOR THE LOT INVOLVED FROM THE LAST 12 MONTHS WAS CONDUCTED AND NO RELATED COMPLAINT(S) WERE FOUND. CONCLUSION: AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY, AND THE REPORTED ISSUE WAS NOT CONFIRMED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING A CATARACT SURGERY, AFTER INJECTING HEALON 5, THE SURGEON STARTED USING ULTRASOUND TO MAKE THE GROOVES. THE SURGEON NOTICED A WHITENING OF THE VISCOUS AND SAW THAT THE CORNEA AT THE INCISION WAS WHITENED AND HAD A CARDBOARD APPEARANCE. THE SURGEON RINSED THE VISCOUS AND USED DUOVISC TO CONTINUE THE SURGERY. AT THE END OF THE SURGERY, SHE HAD TO PUT STITCHES TO KEEP THE INCISION SEALED. IMPORTANT CORNEAL BURN ACCORDING TO HER IMPRESSION AT THE TIME. TO BE FOLLOWED IN POST-OP. THE SURGEON LINKED THE INCIDENT TO THE VISCOUS BUT IT MAY BE A PROBLEM WITH THE IRRIGATION/ASPIRATION. WITHOUT FLOW TO COOL THE TIP, IT CAN QUICKLY HEAT UP. THEY OPERATE WITH SIGNATURE PRO (2 MACHINES IN THIS ACCOUNT) AND WHITESTAR HANDPIECES. THE SURGEON OPERATES IN BIMANUAL (IRRIGATION ON ANOTHER INSTRUMENT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1623163 | HEALON5 PRO OVD | AID, SURGICAL, VISCOELASTIC | LZP | AMO UPPSALA AB | UJ31276 | 05050474701632 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |