FDA Adverse Event Malfunction Summary report: N

BD BACTEC PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 12725036 · Received October 30, 2021

Report

Report Number
2647876-2021-00234
Event Type
Malfunction
Date Received
October 30, 2021
Date of Event
September 30, 2021
Report Date
October 11, 2021
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CUSTOMER REPORTED A POSITIVE ID RESULT FOR BACTEC MEDIA, WHILE USING BIOFIRE FILMARRAY® BLOOD CULTURE IDENTIFICATION BDIC/BCID2 PANELS. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. NO ADDITIONAL TESTING IS REQUIRED AT THIS TIME. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. AS PER PRODUCT INSERT A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBERS OF NONVIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDES, AND SPECIMENS USED FOR INOCULATION. DUE TO THE NATURE OF BIOLOGICAL MATERIALS IN MEDIA PRODUCTS AND INHERENT ORGANISM VARIABILITY, THE USER SHOULD BE COGNIZANT OF POTENTIAL VARIABLE RESULTS IN THE RECOVERY OF CERTAIN MICROORGANISMS. MOLECULAR TESTS PERFORMED ON POSITIVE BLOOD CULTURES WILL DETECT BOTH VIABLE AND NON-VIABLE ORGANISMS COMMONLY FOUND IN CULTURE MEDIA. THEREFORE, MOLECULAR TEST RESULTS SHOULD BE EVALUATED IN CONJUNCTION WITH GRAM STAIN RESULTS IN ACCORDANCE WITH STANDARD-OF-CARE PRACTICES AS WELL AS MANUFACTURER¿S INSTRUCTIONS FOR USE. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) FALSE POSITIVE RESULTS WERE OBTAINED BY THE LABORATORY PERSONNEL. A CULTURE WAS USED TO CONFIRM THE RESULTS AS FALSE POSITIVES. THE CUSTOMER STATED RESULTS WERE REPORTED OUT, BUT THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT BIOFIRE RESULTING AS ACINETOBACTER FOR 442023 BOTTLE, LOT 1160562, BUT IT DOES NOT GROW IN CULTURE"

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1623464 BD BACTEC PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442023 1160562 00382904420239

Patients

Seq Age Sex Outcome Treatment
1