BD SYRINGE 50ML LL BRAZIL
Report
- Report Number
- 1911916-2021-01122
- Event Type
- Malfunction
- Date Received
- October 29, 2021
- Date of Event
- October 1, 2021
- Report Date
- November 16, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVALUATION? YES. D10: RETURNED TO MANUFACTURER ON: 2021-11-15 H6: INVESTIGATION SUMMARY IT WAS REPORTED THERE WAS A BLACK FRAGMENT. TO AID IN THE INVESTIGATION, ONE SAMPLE IN A SEALED PACKAGING BLISTER AND TWO PHOTOS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THE LUER LOK HAS AN EMBEDDED BLACK SPECK. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE TWO PHOTOS PROVIDED SHOW THE SAMPLE RECEIVED. THE EMBEDDED DEGRADED RESIN IN THE COMPONENT TYPICALLY OCCURS AT THE STARTUP OR INTERMITTENTLY DURING THE INJECTION MOLDING PROCESS. THE DEGRADED RESIN CAN BREAK LOOSE AND BE MOLDED INTO COMPONENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 303552, LOT NUMBER 1103477. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. TO MITIGATE THESE ESCAPES, THE FREQUENCY OF INSPECTIONS WERE INCREASED. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.
IT WAS REPORTED THAT BD SYRINGE 50ML LL BRAZIL CONTAINED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRTY SYRINGE (BLACK FRAGMENT).".
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD SYRINGE 50ML LL BRAZIL CONTAINED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DIRTY SYRINGE (BLACK FRAGMENT)."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1618269 | BD SYRINGE 50ML LL BRAZIL | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 1103477 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |