FDA Adverse Event Malfunction Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 12722026 · Received October 29, 2021

Report

Report Number
1221359-2021-03359
Event Type
Malfunction
Date Received
October 29, 2021
Date of Event
October 3, 2021
Report Date
January 12, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT LOT M161492 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-100 / LOT M161492 , TEST BASE PART NUMBER 190-430 / LOT M161492 . THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M161492 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REFERENCE MFR REPORTS: 1221359-2021-03357, 1221359-2021-03358, 1221359-2021-03360.

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M161492 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-100 / LOT M161492 , TEST BASE PART NUMBER 190-430 / LOT M161492 . THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M161492 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REFERENCE MFR REPORTS: 1221359-2021-03357, 1221359-2021-03358, AND 1221359-2021-03360.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. MFR REFERENCE REPORTS: 1221359-2021-03357, 1221359-2021-03358, 1221359-2021-03360.

Description of Event or Problem · 0

THE CUSTOMER REPORTED 5 FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON DIFFERENT DATES. THIS MFR. REPORT ADDRESSES EVENT DATE THREE (3) OF FOUR (4) . THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 TESTED DIRECTLY. (NO VIRAL TRANSPORT MEDIA USED) REPEAT TESTING WAS NOT PERFORMED. RT-PCR CONFIRMATION TESTING WAS PERFORMED WITH COREMAN PLATFORM AND GENERATED NEGATIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1620718 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M161492 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown