FDA Adverse Event Malfunction Summary report: N

ACHIEVE ADVANCE MAPPING CATHETER

MDR report key: 12721737 · Received October 29, 2021

Report

Report Number
9617601-2021-00164
Event Type
Malfunction
Date Received
October 29, 2021
Date of Event
July 6, 2021
Report Date
December 15, 2021
Manufacturer
MEDTRONIC, INC.
Product Code
DRF
PMA / PMN Number
K162892
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UPDATE D1-D4: PRODUCT WAS A 2ACH20 MAPPING CATHETER, NOT A 990063-020 MAPPING CATHETER MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE MAPPING CATHETER 990063-020 WITH AN UNKNOWN LOT NUMBER WAS RETURNED AND ANALYZED. VISUAL INSPECTION OF THE MAPPING CATHETER SHOWED THE DEVICE WAS STUCK INSIDE THE RETURNED BALLOON CATHETER. WHEN TRYING TO RETRIEVE BACK THE MAPPING CATHETER FROM THE BALLOON CATHETER, IT WAS REVEALED THAT THE TIP/LOOP WAS OUT OF SHAPE, KINKED AND RIBBED. IN CONCLUSION, THE MAPPING CATHETER FAILED THE RETURNED PRODUCT INSPECTION DUE TO THE KINKED AND RIBBED LOOP. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

THE MAPPING CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1617343 ACHIEVE ADVANCE MAPPING CATHETER CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT DRF MEDTRONIC, INC. 2ACH20 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Unknown