IMPL TAPERED SCR-V MTX 6M M 5.7MM 11.5MM
Report
- Report Number
- 0002023141-2021-03070
- Event Type
- Injury
- Date Received
- October 29, 2021
- Date of Event
- September 24, 2021
- Report Date
- March 15, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019386
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PERI-IMPLANTITIS IS AN INFECTIOUS DISEASE THAT CAUSES INFLAMMATION OF THE GUM AND THE BONE STRUCTURE AROUND A DENTAL IMPLANT. CHRONIC INFLAMMATION CAUSES BONE LOSS, WHICH CAN LEAD TO IMPLANT FAILURE THUS ITS REMOVAL. INVESTIGATIONS PERFORMED FOR HUNDREDS OF EVENTS WHERE EITHER OF THESE CONDITIONS, INFECTION OR BONE LOSS, OR BOTH, HAVE BEEN REPORTED, HAVE CONCLUDED THAT THE LIKELY CAUSE(S) FOR THE IMPLANT FAILURE IN RELATION TO THESE CONDITIONS ARE EXTERNAL FACTORS. THOSE INCLUDE, MEDICAL CONDITIONS (E.G., DIABETES, BRUXISM, ETC.), PATIENT HABITS (E.G., SMOKING, BAD ORAL HYGIENE ROUTINE) AND USER ERROR (E.G., SURGICAL TECHNIQUE). THERE HAS NOT BEEN ANY INSTANCE WHERE EITHER INFECTION AND/OR BONE LOSS, AND WHEN RIGHT CONDITIONS PREVAIL AND LED TO PERI-IMPLANTITIS WHETHER REPORTED OR NOT ALONG WITH THE OTHER TWO, HAVE RESULTED FROM A MANUFACTURING OR DESIGN DEFECT. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO INFECTION AND RESULT IN BONE LOSS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THE ANALYSIS AND RESULT OF INVESTIGATIONS AND THE ANALYSIS OF PROBABILITY PREVIOUSLY DESCRIBED ARE CONTAINED IN THE SUMMARY INVESTIGATION REPORTS PERFORMED FOR BONE LOSS AND INFECTION, WHICH ARE ATTACHED. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1232401). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP150) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1232401) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FUNCTIONAL : PERI-IMPLANTITIS AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENTS REMAIN NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION. SECTIONS UPDATED: B4: DATE OF THIS REPORT. D9: DEVICE AVAILABILITY AND PRODUCT RETURN DATE. G3: DATE INVESTIGATION RESULTS WERE RECEIVED. G6: TYPE OF REPORT AND FOLLOW UP NUMBER. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED. H6: ADVERSE EVENT PROBLEM CODES.
THERE IS NO UPDATE TO THE ORIGINAL COMPLAINT DESCRIPTION PROVIDED.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ADDITIONAL 510(K) NUMBER IS K013227.
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS. PATIENT PRESENTED WITH PAIN AND A LOOSE IMPLANT. PATIENT WILL RETURN FOR A NEW IMPLANT. TOOTH #30
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1620037 | IMPL TAPERED SCR-V MTX 6M M 5.7MM 11.5MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSV6B11 | 1232401 | 00889024019386 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention |