FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12717219
·
Received October 29, 2021
Report
- Report Number
- 9610877-2021-50199
- Event Type
- Malfunction
- Date Received
- October 29, 2021
- Date of Event
- October 21, 2021
- Report Date
- October 29, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- ODG
- UDI-DI
- 04961333239436
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG38-J10UT-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K200090. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1616031 | PENTAX | DIGITAL LINEAR ULTRASOUND SCOPE | ODG | HOYA CORPORATION PENTAX TOKYO OFFICE | EG38-J10UT | 04961333239436 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|