FDA Adverse Event Other Summary report: N

OPTICAL GUIDANCE PLATFORM

MDR report key: 1271517 · Received December 22, 2008

Report

Report Number
2916710-2007-00011
Event Type
Other
Date Received
December 22, 2008
Date of Event
February 20, 2007
Report Date
February 20, 2007
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
IYE
PMA / PMN Number
K071360
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FLOORSTAND HAS LIMIT SWITCHES, WHICH ARE DESIGNED TO DISABLE GANTRY ROTATION OF THE LINAC FOR CERTAIN BEAM ANGLES WHERE THERE IS A POTENTIAL FOR COLLISION BETWEEN THE GANTRY HEAD AND THE FLOORSTAND AS THE GANTRY MOVES TOWARD THE FLOOR. THESE SWITCHES WERE DESIGNED TO BE COMPATIBLE WITH THE ELEKTA ACCELERATORS (SL 75/5 AND SL 20) AND ARE COMPATIBLE WITH THE VARIAN 2100C, BUT ARE INADEQUATE FOR USE ON THE CLINAC 6-100 AND 600C MODELS - WHICH REQUIRE A BROADER RANGE OF EXCLUSION ANGLES THAT CAN BE SET ON THE FLOORSTAND, AND, POSSIBLY, OTHER THIRD PARTY LINACS AS WELL. NO ADVERSE EVENT HAS BEEN REPORTED, AND THE ISSUE IS NOT DUE TO MALFUNCTION, BUT RATHER USE ON LINACS FOR WHICH THE DEVICE WAS NOT VALIDATED. TWO END USER COMPLAINTS WERE RECEIVED DURING 2007. A PRODUCT NOTIFICATION LETTER WILL BE DISTRIBUTED TO ALL ACTIVE USERS WHO HAVE THE FLOORSTAND INSTALLED WITH A CLINAC 600C OR OTHER NON-VALIDATED MODEL, WITH A DESCRIPTION OF THE PROBLEM AND USER CORRECTIVE ACTION STEPS. ALL CORRECTIVE ACTION WILL BE REPORTED UNDER THE GUIDELINES OF (B) (4). NO FOLLOW-UP REPORTS ARE ANTICIPATED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A POTENTIAL FOR THE GANTRY TO BE DRIVEN INTO THE FLOOR STAND. THE GANTRY LIMIT SWITCHES IN THE FLOOR STAND ARE DESIGNED FOR AN ELEKTA SL75 WITH MOTION INTERRUPTS FITTED FOR 130 DEGREES +/-; WHEN INSTALLING THE LOW ENERGY MACHINES, IT NEEDS TO INTERRUPT AT +/-120 DEGREES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPTICAL GUIDANCE PLATFORM ACCELERATOR, LINEAR, MEDICAL IYE VARIAN MEDICAL SYSTEMS HZ5

Patients

Seq Age Sex Outcome Treatment
1