PROGRESSA FRAME
Report
- Report Number
- 1824206-2021-00514
- Event Type
- Malfunction
- Date Received
- October 28, 2021
- Date of Event
- October 22, 2021
- Report Date
- October 28, 2021
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE HILLROM TECHNICIAN FOUND THE EXTERNAL ALARM NEEDED TO BE REPLACED. THE EXTERNAL ALARM IS THE SOURCE OF THE BRAKE NOT SET ALARM. PER THE PROGRESSA USER MANUAL: HE BRAKE NOT SET IS AN AUDIBLE AND VISUAL ALARM. THE ALARM WILL SOUND, AND A MESSAGE WILL SHOW ON THE GCI, WHEN THE BED IS CONNECTED TO AN AC POWER SOURCE AND THE BRAKE IS NOT SET. WHEN THE BED IS CONNECTED TO AC POWER AND THE BRAKES ARE NOT SET, AN ALARM SOUNDS AND A MESSAGE SHOWS ON THE GCI. WHEN AC POWER IS REMOVED, THE ALARM WILL STOP AND THE GCI WILL TURN OFF. A SEARCH OF THE HILLROM MAINTENANCE RECORDS DID NOT SHOW HILLROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE EXTERNAL ALARM TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE BED HAD NO AUDIBLE ALARM. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1611812 | PROGRESSA FRAME | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | HILL-ROM BATESVILLE | P7500A000131 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |