FDA Adverse Event Malfunction Summary report: N

SUPERCORE BIOPSY INSTRUMENT 18GA X 6CM

MDR report key: 12712501 · Received October 28, 2021

Report

Report Number
0001625425-2021-01110
Event Type
Malfunction
Date Received
October 28, 2021
Date of Event
October 1, 2021
Report Date
November 19, 2021
Manufacturer
ARGON MEDICAL DEVICES
Product Code
KNW
UDI-DI
00886333005963
PMA / PMN Number
K974814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE UNOPENED DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION FOUND THAT THE BODY OF THE DEVICE WAS STARTING TO SPLIT PROXIMALLY, AND THE SPRING RELEASED. THEREBY CONFIRMING THE COMPLAINT. ARGON MEDICAL DEVICES HAS RECEIVED OTHER COMPLAINTS THAT THE SUPERCORE SEMI-AUTOMATIC BIOPSY INSTRUMENT IS COMING APART DURING OR PRIOR TO USE. ARGON HAS CONDUCTED AN INTERNAL INVESTIGATION AND TRACKED THE AFFECTED PARTS TO A NARROW TIME FRAME RESULTING FROM A SPECIFIC MANUFACTURING EVENT. THE PLASTIC HOUSING AND PLUNGER CAN BE SEPARATED MORE EASILY THAN NORMAL FOR THE LOTS MANUFACTURED DURING THIS TIME FRAME. CAPA 2021-061 HAS BEEN OPENED TO DOCUMENT OUR INVESTIGATION INTO THE CAUSE OF THIS PROBLEM AND THE CORRECTIVE ACTION THAT ARE BEING TAKEN TO ENSURE THAT THIS NEVER HAPPENS AGAIN. TO ENSURE CONTINUED CUSTOMER SATISFACTION, ARGON MEDICAL DEVICES HAS DECIDED TO ISSUE A VOLUNTARY RECALL OF THE AFFECTED LOTS BECAUSE OF THE HIGH RATE OF REPORTS OF UNINTENTIONAL DISASSEMBLY OF THESE DEVICES.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED, AND EVALUATION IS ANTICIPATED BUT HAD NOT YET BEGUN. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE DEVICE HAS BEEN EVALUATED.

Description of Event or Problem · 0

SPRING POPPED OPEN AS THE NEEDLE WAS PLACED WITHIN THE NECK @ THYROID NODULE.

Description of Event or Problem · 0

SPRING POPPED OPEN AS THE NEEDLE WAS PLACED WITHIN THE NECK @ THYROID NODULE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1614601 SUPERCORE BIOPSY INSTRUMENT 18GA X 6CM SUPERCORE KNW ARGON MEDICAL DEVICES 701118060 11374235 00886333005963

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other