SUPERCORE BIOPSY INSTRUMENT 18GA X 6CM
Report
- Report Number
- 0001625425-2021-01110
- Event Type
- Malfunction
- Date Received
- October 28, 2021
- Date of Event
- October 1, 2021
- Report Date
- November 19, 2021
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- KNW
- UDI-DI
- 00886333005963
- PMA / PMN Number
- K974814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ONE UNOPENED DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION FOUND THAT THE BODY OF THE DEVICE WAS STARTING TO SPLIT PROXIMALLY, AND THE SPRING RELEASED. THEREBY CONFIRMING THE COMPLAINT. ARGON MEDICAL DEVICES HAS RECEIVED OTHER COMPLAINTS THAT THE SUPERCORE SEMI-AUTOMATIC BIOPSY INSTRUMENT IS COMING APART DURING OR PRIOR TO USE. ARGON HAS CONDUCTED AN INTERNAL INVESTIGATION AND TRACKED THE AFFECTED PARTS TO A NARROW TIME FRAME RESULTING FROM A SPECIFIC MANUFACTURING EVENT. THE PLASTIC HOUSING AND PLUNGER CAN BE SEPARATED MORE EASILY THAN NORMAL FOR THE LOTS MANUFACTURED DURING THIS TIME FRAME. CAPA 2021-061 HAS BEEN OPENED TO DOCUMENT OUR INVESTIGATION INTO THE CAUSE OF THIS PROBLEM AND THE CORRECTIVE ACTION THAT ARE BEING TAKEN TO ENSURE THAT THIS NEVER HAPPENS AGAIN. TO ENSURE CONTINUED CUSTOMER SATISFACTION, ARGON MEDICAL DEVICES HAS DECIDED TO ISSUE A VOLUNTARY RECALL OF THE AFFECTED LOTS BECAUSE OF THE HIGH RATE OF REPORTS OF UNINTENTIONAL DISASSEMBLY OF THESE DEVICES.
THE DEVICE WAS RETURNED, AND EVALUATION IS ANTICIPATED BUT HAD NOT YET BEGUN. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE DEVICE HAS BEEN EVALUATED.
SPRING POPPED OPEN AS THE NEEDLE WAS PLACED WITHIN THE NECK @ THYROID NODULE.
SPRING POPPED OPEN AS THE NEEDLE WAS PLACED WITHIN THE NECK @ THYROID NODULE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1614601 | SUPERCORE BIOPSY INSTRUMENT 18GA X 6CM | SUPERCORE | KNW | ARGON MEDICAL DEVICES | 701118060 | 11374235 | 00886333005963 |