FDA Adverse Event Injury Summary report: N

GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM

MDR report key: 12712104 · Received October 28, 2021

Report

Report Number
2024168-2021-09692
Event Type
Injury
Date Received
October 28, 2021
Date of Event
September 27, 2021
Report Date
January 7, 2022
Manufacturer
ABBOTT VASCULAR
Product Code
MAF
PMA / PMN Number
H000001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A VISUAL AND DIMENSIONAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE TESTED AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE INVESTIGATION DETERMINED THE REPORTED FAILURE TO ADVANCE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH A FREE PERFORATION WITH EXTRAVASATION IN THE LEFT CIRCUMFLEX ARTERY. A 3.5X26MM RX GRAFTMASTER COVERED STENT FAILED TO CROSS DUE TO ANATOMY. A DRUG-ELUTING STENT WAS USED TO SEAL THE PERFORATION AND COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1610922 GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM CORONARY STENT DELIVERY SYSTEM MAF ABBOTT VASCULAR 1052441

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other