FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12708876 · Received October 27, 2021

Report

Report Number
1221359-2021-03337
Event Type
Malfunction
Date Received
October 27, 2021
Report Date
December 3, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M160956 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT M160956 AND TEST BASE PART NUMBER 190-430 / LOT M160956. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M160956 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY ON A DIRECT TESTED NASOPHARYNGEAL PURITAN DRY SWAB. CONFIRMATION TESTING WITH PCR GENERATED NEGATIVE RESULTS. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1604932 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19, QJR ABBOTT DIAGNOSTICS SCABOROUGH M160956 10811877011269

Patients

Seq Age Sex Outcome Treatment
1 Unknown