INFUSOMAT®
Report
- Report Number
- 9610825-2021-00444
- Event Type
- Death
- Date Received
- October 27, 2021
- Date of Event
- September 23, 2021
- Report Date
- March 1, 2022
- Manufacturer
- B. BRAUN MELSUNGEN AG
- Product Code
- FRN
- UDI-DI
- 04046964660887
- PMA / PMN Number
- K142596
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL, INC. INTERNAL REPORT (B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS INSPECTED AND TESTED AND THE REPORTED ISSUE WAS NOT PRESENT DURING THE INVESTIGATION. THE LOG REVIEW SHOWS CONTINUING INFUSIONS FROM (B)(6) 2021 AND AT 11:07AM A NEW RATE SET TO 54.45ML/HR (DL: DOB500), THIS CONTINUED THROUGH THE TIME OF INCIDENT 2:33-3:45PM ON (B)(6) 2021 WITH AN AIR BUBBLE ALARM OCCURRING AT 3:46PM AND AN INFUSION START AT 3:49PM. INFUSIONS CONTINUED AND WERE STOPPED AT 5:59PM. THE LOG SHOWS NO DEVICE ALARMS ON (B)(6) 2021. NOTE: MAINS POWER OFF AT (B)(6) 2021 AND 2:32PM AND ON AT 4:06PM. ABOVE TIMES ARE ACTUAL; LOG SHOWS PUMP TIME 1HR 18MIN LATER THAN ACTUAL TIME. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE A FOLLOW-UP REPORT WILL BE FILED.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER (B)(4). PUMP LOG REVIEW INDICATED THAT THE PUMP (611626) INFUSED CONTINUOUSLY DURING TRANSPORT AND ALARMED FOR "AIR BUBBLE" AFTER ARRIVAL TO THE ICU AT 15:46. THE ALARM WAS ACKNOWLEDGED BY THE USER AND THE INFUSIONS WERE RESTARTED. THE DATA SUPPORTS THAT THE PUMPS OPERATED AS INTENDED AND THERE IS NO INDICATION OF ANY PUMP MALFUNCTION THAT CONTRIBUTED TO THE PATIENT OUTCOME AT THIS TIME. THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.
AS REPORTED BY THE USER FACILITY: THE PATIENT WAS TAKEN TO THE CATH LAB EMERGENTLY ON (B)(6) 2021 AT 1433 (CST), PER CHART. PATIENT TRANSPORTED FROM CATH LAB TO ICU AT 1532 (CST). THE CATH LAB IS ON THE SECOND FLOOR AND THE ICU SHE WAS GOING TO IS ON THE 3RD FLOOR. PER THE CHART, SHE ARRIVED TO ICU AT 1545 WITH NO PULSE, NO BP AND THE NURSE NOTICED THE PUMPS WERE ALL ALARMING AND PAUSED. SHE EXPIRED AT 2117 ON (B)(6) 2021 (CST). WE ARE TRYING TO DETERMINE WHETHER THE PUMPS WERE MANUALLY PAUSED, AIR IN THE LINES, OR WHAT IN THE 13 MINUTES IT TOOK TO TRANSPORT HER FROM ONE AREA TO ANOTHER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1600131 | INFUSOMAT® | PUMP, INFUSION | FRN | B. BRAUN MELSUNGEN AG | 8713051U | 96990691I9 | 04046964660887 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Death |