SIL. COATED I/A 45° MILL
Report
- Report Number
- 0001920664-2021-00142
- Event Type
- Injury
- Date Received
- October 27, 2021
- Date of Event
- October 5, 2021
- Report Date
- October 5, 2021
- Manufacturer
- BAUSCH + LOMB GMBH
- Product Code
- HNN
- PMA / PMN Number
- K073023
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE EVALUATION HAS BEEN COMPLETED. ONE OPENED 85910S I/A HANDPIECE FROM LOT FS21037856 WAS RETURNED. THE EXPIRATION DATE ON THE LABEL IS 2024-02. VISUAL INSPECTION FOUND PARTICULATES AND DRIED SOLUTIONS VISIBLE ON THE IRRIGATION SLEEVE. MICROSCOPIC EXAMINATION WAS PERFORMED AND FOUND NO DEFECTS THAT WOULD CAUSE EXPOSURE TO A SHARP EDGE. A FUNCTIONAL TEST WAS PERFORMED AND FOUND THAT THE ASSEMBLY PERFORMED AS INTENDED. THE ROOT CAUSE OF THE COMPLAINT COULD NOT BE DETERMINED. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. CAPSULE RAPTURE IS A WELL-KNOWN RISK OF THE SURGERY. USE OF AN INSTRUMENT WITH A SILICONE SLEEVE REDUCES THIS RISK BUT DOES NOT COMPLETELY ELIMINATE IT. IN LAST 2 YEARS THIS IS THE ONLY RECORDED CASE OF CAPSULE DAMAGE FOR OVER 1,6 MLN UNITS WITH A SILICONE, SOLD BY BAUSCH AND LOMB IN LAST 2 YEARS. THE SURGEON PLACED A THREE PIECE LENS IN THE PATIENT AND THE PATIENT IS DOING WELL.
THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DEVIATIONS OR ISSUES WERE FOUND. THE LOT WAS MANUFACTURED ACCORDING TO SPECIFICATION. THIS INVESTIGATION IS ONGOING.
THE SURGEON REPORTED WHILE POLISHING THE IA TIP BROKE THE CAPSULE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1601873 | SIL. COATED I/A 45° MILL | HANDHELD SURGICAL INSTRUMENT | HNN | BAUSCH + LOMB GMBH | 85910S | 222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |