FDA Adverse Event Injury Summary report: N

SIL. COATED I/A 45° MILL

MDR report key: 12706449 · Received October 27, 2021

Report

Report Number
0001920664-2021-00142
Event Type
Injury
Date Received
October 27, 2021
Date of Event
October 5, 2021
Report Date
October 5, 2021
Manufacturer
BAUSCH + LOMB GMBH
Product Code
HNN
PMA / PMN Number
K073023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE EVALUATION HAS BEEN COMPLETED. ONE OPENED 85910S I/A HANDPIECE FROM LOT FS21037856 WAS RETURNED. THE EXPIRATION DATE ON THE LABEL IS 2024-02. VISUAL INSPECTION FOUND PARTICULATES AND DRIED SOLUTIONS VISIBLE ON THE IRRIGATION SLEEVE. MICROSCOPIC EXAMINATION WAS PERFORMED AND FOUND NO DEFECTS THAT WOULD CAUSE EXPOSURE TO A SHARP EDGE. A FUNCTIONAL TEST WAS PERFORMED AND FOUND THAT THE ASSEMBLY PERFORMED AS INTENDED. THE ROOT CAUSE OF THE COMPLAINT COULD NOT BE DETERMINED. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. CAPSULE RAPTURE IS A WELL-KNOWN RISK OF THE SURGERY. USE OF AN INSTRUMENT WITH A SILICONE SLEEVE REDUCES THIS RISK BUT DOES NOT COMPLETELY ELIMINATE IT. IN LAST 2 YEARS THIS IS THE ONLY RECORDED CASE OF CAPSULE DAMAGE FOR OVER 1,6 MLN UNITS WITH A SILICONE, SOLD BY BAUSCH AND LOMB IN LAST 2 YEARS. THE SURGEON PLACED A THREE PIECE LENS IN THE PATIENT AND THE PATIENT IS DOING WELL.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DEVIATIONS OR ISSUES WERE FOUND. THE LOT WAS MANUFACTURED ACCORDING TO SPECIFICATION. THIS INVESTIGATION IS ONGOING.

Description of Event or Problem · 1

THE SURGEON REPORTED WHILE POLISHING THE IA TIP BROKE THE CAPSULE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1601873 SIL. COATED I/A 45° MILL HANDHELD SURGICAL INSTRUMENT HNN BAUSCH + LOMB GMBH 85910S 222

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention