FDA Adverse Event Malfunction Summary report: N

PEGASUS YEL 24GA X 0.75IN QSYTE-CAP Y

MDR report key: 12703000 · Received October 27, 2021

Report

Report Number
8041187-2021-00929
Event Type
Malfunction
Date Received
October 27, 2021
Date of Event
September 28, 2021
Report Date
November 30, 2021
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
PMA / PMN Number
PENDING
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1153670. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ALTHOUGH A VIDEO WAS SUBMITTED FOR EVALUATION, THE AFFECTED SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT PEGASUS YEL 24GA X 0.75IN QSYTE-CAP Y LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DURING PERIPHERAL INFUSION FOR THE PATIENT, FLUID LEAKAGE WAS FOUND FROM THE MOUTH OF THE WHITE END CAP."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PEGASUS YEL 24GA X 0.75IN QSYTE-CAP Y LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DURING PERIPHERAL INFUSION FOR THE PATIENT, FLUID LEAKAGE WAS FOUND FROM THE MOUTH OF THE WHITE END CAP."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1604133 PEGASUS YEL 24GA X 0.75IN QSYTE-CAP Y INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 1153670

Patients

Seq Age Sex Outcome Treatment
1 Unknown